Configuring and operating the IRG 2000
Syscompact 2000 M pro
44 / 98
822-175-2
9.4 Adjusting the colours of the display elements
You can adjust the colours used to display the traces, the cursors and the background of the
reflection image.
>
System
>
Colors
1. Press the
MENU
key.
2. Use the rotary knob to select the
System
menu item and press the rotary knob.
3. Use the rotary knob to select the
Colors
menu item and press the rotary knob.
4. Use the rotary knob to adjust the colours for the various display elements.
5. To save the settings, turn the rotary knob until the
OK
button is selected, then press the
rotary knob.
9.5 Selecting the system of units
>
System
>
Unit system
1. Press the
MENU
key.
2. Use the rotary knob to select the
System
menu item and press the rotary knob.
3. Use the rotary knob to select the
Unit system
menu item and press the rotary knob.
4. Select the system of units to be used to display the distance and speed information:
Metric
to display in metres
English
to display in feet
9.6 Evaluating reflection images with cursors
Cursors enable you to mark fault positions precisely or to measure distances. There are 3
cursors available: Zero cursor (
Cursor # 0
),
Cursor # 1
and
Cursor # 2
.
The distances of Cursors 1 and 2 are calculated and displayed automatically as follows:
Distance between Cursor 1 and the zero cursor
Distance between Cursor 2 and the zero cursor.
9.6.1 Setting cursors
1. Perform a measurement.
Further information: Chapter
Cable fault pre-location
(on page 52)
2. Press the
CURSOR
key.
3. Turn the rotary knob until the desired cursor is selected, then press the rotary knob.
The button has a green background.
4. Use the rotary knob to position the cursor and press the rotary knob.
The button has an orange background.
5. To save the position of the cursor, adjust the position of the cursor if necessary and press
the rotary knob again.
The cursor position is saved.