EMD-RMTR.. User’s Manual
19
6.2.28 FwdTotal Lo, hi
Posi ve total volume high byte and low byte can change forthcoming and reverse total value, and be used to
maintenance and instead.
User use 5 byte code to enter, and can modify the posi ve accumula ng volume (∑+). Usually, it is unsuitable
to exceed the maximum the counter set(999999999).
6.2.29 RevTotal Lo
、
hi
User use 5 byte code to enter, and can modify the nega ve accumula ng volume (∑-). Usually, it is unsuitable
to exceed the minimum the counter set(999999999).
6.2.30 PlsntLmtEn
For paper pulp, slurry and other serosity, the flow measure will have "cuspidal disturb", because the solid grain
friction or concussion the measure electrode. -RM converters
use varia on restrain arithme c to conquer the disturbing by designing three parameters to select disturb
character.
Set it "enable", start varia on restrain arithmetic; set it "disable", close varia on restrain arithmetic.
6.2.31PlsntLmtVl
This coefficient can disturb the varia on of cuspidal disturb, and calculate as percent of flow velocity, thus ten
grades: 0.010m/s, 0.020m/s, 0030m/s, 0.050m/s, 0.080m/s, 0.100m/s, 0.200m/s, 0.300m/s, 0.500m/s,
0.800m/s, and the smaller percent, the higher delicacy of cuspidal restrain.
Note: when using it, must test for select by the fact, and some mes it is not the higher delicacy is good.
6.2.32 PlsntDelay
This coefficient can select the width of me of restrain cuspidal disturb and the unit is ms. If the dura on is
shorter than flow change in some me, –RM series will think it is cuspidal disturb, and if it is longer, –RM series
will think it is natural. It also needs to select parameter in fact.
6.2.33User’s password 1~4
Users can use 5 grades of passwords to correct these passwords.
6.2.34 Analog Zero
When the converters are made in the factory, output current has been calibrated to zero scale, that is,
accurate 0mA or 4mA output.
6.2.35 Anlg Range
When the converters is made in the factory, output current have been calibrated to full scale, that is, accurate
10mA or 20mA output.
6.2.36Meter Fact
This fact is the special one of sensor-made-factory and the factory use this fact to unite –RM series
electromagne c flowmeters converters to make sure all the instruments can interchange by 0.1%.