Operation manual
55
-5V power supply
IU2?
Response contains measured value of -5V internal power supply. Accuracy is not guaranteed (value is
informative).
Example:
If query „IU2?<cr>” is sent, flowmeter returns response in format „4.994<cr>“ for voltage 4.994 V.
+15V power supply
IU3?
Response contains measured value of +15V internal power supply. Accuracy is not guaranteed (value is
informative).
Example:
If query „IU3?<cr>” is sent, flowmeter returns response in format „14.84<cr>“ for voltage 14.84 V.
Coils voltage
IU4?
Response contains measured value of voltage on coils. Accuracy is not guaranteed (value is informative).
Example:
If query „IU4?<cr>” is sent, flowmeter returns response in format „10.12<cr>“ for voltage 10.12 V.
+2.5V power supply
IU5?
Response contains measured value of +2.5V internal power supply. Accuracy is not guaranteed (value is
informative).
Example:
If query „IU5?<cr>” is sent, flowmeter returns response in format „2.55<cr>“ for voltage 2.55 V.
Coils current
IU6?
Response contains measured value of coils current (mA). Accuracy is not guaranteed (value is informative).
Example:
If query „IU6?<cr>” is sent, flowmeter returns response in format „99.85<cr>“ for current 99.85 mA.
Excitation coils resistance
ICO?
Response contains measured value of excitation coil resistance. Accuracy is not guaranteed (value is
informative).
Example:
If query „ICO?<cr>” is sent, flowmeter returns response in format „104.7<cr>“ for resistance 104.7 Ω