![BASI BFM921 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/basi/bfm921/bfm921_user-manual_477905048.webp)
Electromagnetic Flowmeter
BF
M921
BASI
48
Operation manual
Auxiliary volume counter Reset
CLRAV
Command resets “Auxiliary volume counter”.
BF
M921 confirms execution with string „Ok<cr>”.
Example:
Command „CLRAV<cr>” resets the
Auxiliary volume counter
.
Timed volume counter Reset
CLRTV
Command resets “Timed volume counter”.
M921 confirms execution with string „Ok<cr>”.
Example:
Command „CLRTV<cr>” resets the
Timed volume counter
.
5.2.2.6
Datalogger parameters
Datalogger step
DST(?)<CPD> { 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 }
Datalogger can be set:
0
Datalogger is Off
1
Datalogger sampling rate is 5 minutes.
2
Datalogger sampling rate is 10 minutes.
3
Datalogger sampling rate is 15 minutes.
4
Datalogger sampling rate is 30 minutes.
5
Datalogger sampling rate is 45 minutes.
6
Datalogger sampling rate is 60 minutes.
7
Datalogger sampling rate is 120 minutes.
8
Datalogger sampling rate is 180 minutes.
9
Datalogger sampling rate is 240 minutes.
M921 confirms datalogger step with string „Ok<cr>”.
Example RS232:
Command „DST0<cr>” switches datalogger “Off”. If query „DST?<cr>” is sent, flowmeter returns response
in format „0<cr>”.
Datalogger number of samples
DNR?
Response contains number of flowrate samples stored in datalogger.
Example:
If query „DNR?<cr>” is sent, flowmeter returns response in format „252<cr>“ for 252 samples in datalogger.
Datalogger filling (percentage)
DPC?
Response contains datalogger filling in percent.
Example:
If query „DPC?<cr>” is sent, flowmeter returns response in format „14<cr>“ for 14% datalogger full.