IFX-21 User Guide
BASF
Maintenance
7/12/2019
V004
14 of 22
A
NALOG
O
UTPUT
O
PERATION
The analog output value will depend on the status of the full setup and connections.
With the loop powered, the output value could be the following:
IFX Power
Pyro Connection
Output Value
Yes/No
Yes/No
≤1.8mA, Analog Output malfunction or failure
No
Yes/No
2.2mA, Default Output
Yes
Yes
2.7mA, Analog Output is NOT enabled in Pyrometer settings
Yes
No
“Failure signal (mA)” [Default 3.6mA]
Yes
Yes
Measured Temperature Reading
A
NALOG
O
UTPUT
C
ONFIGURATION
Use TemperaSure to configure the analog output scales.
6. M
AINTENANCE
Periodically check device for signs of water, oil, dirt, etc. If necessary, clean it by
disconnecting the DC power and wiping with a soft, slightly damp cloth.