BARTINGTON INSTRUMENTS
Page 39 of 82 OM0408/49
4. Place the probe in firm contact with the material to be measured and take a reading.
Note:
For materials with susceptibility values less than 5-10 CGS (10 on the LCD readout),
it will usually be necessary to zero the instrument between each reading.
12. MS2G Single Frequency Sensor
Figure 15. MS2G single frequency sensor.
Key to Figure 15
1. TNC coaxial cable connector
2. Calibration sample
3. Height setting screw
4. Sample measurement cavity
5. Retaining nut
12.1. General Description
The MS2G sensor type is intended for use with a 1cc volume sample vial. It comprises a metal
enclosure containing the electronic circuitry, at the end of which is mounted a rectangular block
containing the measurement cavity.
Connection to the MS2 meter is via the standard TNC-TNC coaxial cable. Samples are inserted
into the top of the cavity. An adjustable mechanical stop can be set to centralise the sample
material within the measurement zone. This permits a sample volume down to 0.2cc to be used.
Markings on the front of the cavity show the position of the measurement zone. The sensor
operates at a single frequency. A calibration check sample is provided.
4
3
2
5
1