background image

BARTINGTON INSTRUMENTS

 

Page 15 of 20 OM0382/9

6. Operating Instructions 

Refer to figures 1, 4 and 5.

6.1. Switching On

Connect the probe selected to the probe socket and switch on the magnetometer.

Note: 

The plug locking mechanism requires that the outer shell of the plug is retracted 

(towards cable) during insertion and removal.

For the first 2-5 seconds following switch-on, the battery voltage will be displayed on the LCD. 

Note: 

If the battery voltage is below 5V then the instrument must be put on charge 

immediately. An audible alarm will also signal when the battery requires recharging. The 
instrument should never be left in a discharged state as the capacity of the battery will be 
adversely affected.

Some increase in noise can be expected during the first few minutes of charging. The charging 
circuit is temperature compensated and the instrument may be left on charge indefinitely 
without harm to the battery.

The optimum stability of the magnetometer will be achieved in approximately 10 seconds 
following switch-on. Check that the instrument is working by subjecting the probe to a slowly 
varying magnetic field. This is best done for a free probe by rotating it in the earth’s magnetic 
field and observing the change in values. The geomagnetic field will be between 30

m

T and 90

m

T. 

6.2. Magnetic Hygiene

Any object that is magnetic, or has a high magnetic susceptibility, may cause a change in 
the surrounding magnetic field and therefore to the field measured by the Mag-01H. Ensure 
that the operator and the surrounding environment are magnetically clean when perfoming 
measurements.

Caution: 

Avoid any exposure of the probes to very high fields. (Fields up to 1T have been 

shown to cause no permanent damage or impaired performance.) The instrument will fully 
recover from severe overload in around 5 seconds.

6.3. Using the Offset Control (Mag-01H only)

To obtain the maximum resolution on the autoranging LCD, the bipolar offset control  can be 
used to add a positive or negative offset to the field strength experienced by the probe.

Example:

 if F=90.00

μ

T and offset is set to -80, then the display will then show +10.00.

Содержание Mag-01

Страница 1: ...Operation Manual for Mag 01 Mag 01H Single Axis Fluxgate Magnetometers with Mag Probes B G...

Страница 2: ...7 4 1 4 Operation 7 4 2 Circuitry 8 4 2 1 Additional Circuitry for Mag 01H 9 4 3 Power Supplies and Battery Charging 10 4 4 The Autoranging Display 11 4 5 Analogue Output 12 5 The Probes 12 5 1 Intro...

Страница 3: ...ment 16 6 6 Electromagnetic Compatibility 16 7 Troubleshooting Care and Maintenance 17 7 1 Troubleshooting 17 7 2 Checking the System 17 7 2 1 Battery Check 17 7 2 2 Functional Check 17 7 2 3 Over ran...

Страница 4: ...njury or damage to your instrument or both could result if the caution is ignored Identifies items that must be disposed of safely to prevent unnecessary damage to the environment Note Provides useful...

Страница 5: ...gure 1 1 Power switch 1 Waterproof connection to probe 1 Charge light illuminated when internal batteries are on charge 1 4 digit LCD display Displays directly in microtesla Autoranging decimal point...

Страница 6: ...of connection to probe 9 Charge light illuminated when internal batteries on charge 10 Offset toggle switch See 5 11 4 digit LCD display Displays directly in microtesla at x1 sensitivity See Using the...

Страница 7: ...voltage for connection to a DVM or acquisition unit A rechargeable battery is included making the unit suitable for portable use The use of low signature components including a non magnetic battery al...

Страница 8: ...DEMODULAT OR CHARGE INDICAT OR CONSTA NT VOLT A GE BAT TERYCHARGER 12V d c INLET 19 999 A D CONVERTOR 4 1 2 DIGIT LCD ANALOG OUTPUT OUTPUT FILT ER 3dB 10Hz 0v X10 AT TENUAT OR RANGE CONTROL SW ITCH S...

Страница 9: ...fed back to the probe via a current converter and the feedback winding where nulling of the magnetic field occurs In this way with the error removed a precise measurement of field strength is obtained...

Страница 10: ...urs and overcharging is virtually impossible The 12V 6 18V DC external power supply inlet is protected against incorrect polarity supplies A charge indicator on the front panel is illuminated when the...

Страница 11: ...ctive material to form an electrical pathway between the positive and negative terminals of the battery iv Replace the instrument top cover Note To maintain the life span of the internal rechargeable...

Страница 12: ...2V range monitored by the LCD and will be 5V full scale Due to the internal resistance of the high field probes the analogue output voltage can only be considered linear with respect to the magnetic...

Страница 13: ...as indicated in the product brochure Each is identified by a code letter on the label affixed close to the plug on the probe cable For axial construction the sensing direction is in the direction of t...

Страница 14: ...ostat This length can be extended to a maximum of 5m if required Note The cryogenic probe can be mounted by bonding using a minimal mount epoxy adhesive or by loosely supporting the probe within a tub...

Страница 15: ...will be achieved in approximately 10 seconds following switch on Check that the instrument is working by subjecting the probe to a slowly varying magnetic field This is best done for a free probe by r...

Страница 16: ...Supplies and Battery Charging Remove the two case retaining screws in the base of the instrument and remove the top cover Disconnect the battery leads Mag 01 or the battery connector Mag 01H and fit...

Страница 17: ...the magnetometer The battery voltage is shown for a few seconds at switch on and should be between 6 5V and 7V Disconnect the battery charger Switch the magnetometer off and on again The battery volta...

Страница 18: ...accordance with local regulations Refer to the IATA website www iata org for regulations regarding air transport Caution If there is any doubt at all as to the integrity of a battery for example crac...

Страница 19: ...BARTINGTON INSTRUMENTS Page 19 of 20 OM0382 9 Notes...

Страница 20: ...ton Instruments Limited in the following countries United Kingdom Australia Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thor...

Отзывы: