43
7.
Pull the ultrafilter out of the unit and turn it to the
left side. (This will make it easier to disconnect
the bottom fitting.)
8.
Using the tool supplied with the ultrafilter, simul-
taneously push and hold the collet on the ultrafil-
ter in while pulling the bottom fitting out of the
connector. See photo.
9.
Install new ultrafilter according to “Ultrafilter
Installation” in the “Installation” section of this
manual.
10. Sanitize the NANOpure DIamond Life Science
(UV/UF) according to the “System Sanitization”
section of this manual.
M
AINTENANCE AND
S
ERVICING
Ultrafilter Installation
Metal bracket
Velcro strap
Ultrafilter Connections - Right side view
Bottom connector
Top connector
Tool used for ultrafilter removal
Collet
Collet
Tubing with
luer fitting
Collet