56-101-01175 Revision P Page | 38
Figure 43
Test Set Configuration, Updated Altitude Correction
4.
Control Mode:
Allows user to select the default: 2-Channel (Dual Ps, Pt) mode or
1-Channel (Pt
Only) mode. You will see the operating “Mode” is always indicated in
the status window located on the top right-side of the UI display (refer to Figure 44).
5. The Pt Only mode can be used to drive an Airspeed or MACH indicator that has its Static
(Ps) port vented to ambient (
local barometric pressure
). When the single channel Pt Only
mode is selected, the Test Set will perform a Go To Ground function before transitioning
into this mode to ensure the Test Set Static (Ps) port is at ambient barometric pressure.
At the conclusion of the Go To Ground the Test Set will leave the Static (Ps) port vented
to ambient while driving the Pitot (Pt) port to the Target pressure specified by the user.
Note
: The DPS1000 always powers up into the default: 2-Channel (Dual Ps, Pt) mode.
Figure 44
Setup 1: Control Mode(s)
6.
Copy History file to USB:
This allows the user to copy and transfer the history log to
an external USB flash drive. Once on the flash drive, the file can be opened and
reviewed using any tool that allows viewing of ascii text files. It is recommended to
email this file to Barfield to aide in trouble shooting some issues/problems when the
need occurs. In case of required technical support, Barfield personnel may request
that this file be forwarded to them for troubleshooting purposes.