DALT55 INSTRUCTION SUPPLEMENT
61-101-02184_02186-B Page
10/14
Jan /10 /2013
(c) Cal Coeffs.
Displays linearization coefficients:
Offset pressure (inHg)
Slope
linearization
Date last calibration was performed
(d) Config.
This option allows user to configure instrument, selecting:
Stand-alone:
DALT55
is not communicating with
DAS650
.
Two-instrument:
DALT55
is communicating with
DAS650
.
Note:
In the two-instrument configuration, the serial link
indicator shows at bottom of Aeronautical, Pressure,
Leak Test and EPR Test displays. The indicator flashes
when link is not functional.
Press
CHANGE
; then
SAVE
to change configuration.
Level 2
maintenance submenu is password-protected. User
accesses calibration procedure, sets date/time and changes
password and enables automatic power off.
NOTE: The initial default password is “111”.
CHART 3 LEVEL 2 MAINTENANCE SUBMENU
(a) Calibration
User may perform calibration and change the units system for
calibration by pressing the UNITS key. Continue pressing
UNITS key to “scroll” to m (meters), inHg, mbar, Psi, or ft
screens. Press ENTER to store.
LEVEL 2
CALIBRATION
PASSWORD
PASSWORD
DATE / TIME
DATE
TIME
Enter new password
Confirm new password
AUTO OFF