Product Description

PowerMidline™ Catheters are a family of peripherally placed catheters made from radi-
opaque body-softening polyurethane materials. Each PowerMidline™ Catheter is designed 

with kink-resistant, reverse taper design. Catheters are packaged in a tray with accessories 

for reliable short term (less than 30 days) vascular access. The PowerMidline™ Catheters 

are suitable for use with power-injectors.

Indications

The PowerMidline™ Catheters are indicated for short term access to the peripheral venous 
system for selected intravenous therapies, blood sampling, and power injection of contrast 

media. These catheters may be used for any patient population with consideration given to 

adequacy of vascular anatomy and appropriateness of the procedure. The PowerMidline™ 

Catheters are suitable for use with power injectors.

Contraindications

The device is contraindicated whenever:

•  The presence of device-related infection, bacteremia, or septicemia is known or sus-

pected.

• 

The patient’s body size is insufficient to accommodate the size of the implanted device.

•  The patient is known or is suspected to be allergic to materials contained in the device.
•  Local tissue factors and/or past treatment will prevent proper device stabilization and/

or access.

Warnings

General Warnings

•  When using alcohol or alcohol-containing antiseptics with polyurethane midlines, care 

should be taken to avoid prolonged or excessive contact. Solutions should be allowed 
to completely dry before applying an occlusive dressing. Chlorhexidine gluconate and/

or povidone iodine are the suggested antiseptics to use.

•  Alcohol should not be used to lock, soak or de-clot polyurethane midlines because alco-

hol is known to degrade polyurethane catheters over time with repeated and prolonged 
exposure.

•  Acetone-based solution and polyethlene glycol containing ointments should not be used 

with polyurethane catheters, as these may cause failure of the device.

•  Therapies not appropriate for midline catheters include those therapies requiring central 

venous access. Refer to standards of practice and institutional policies.

•  Do not wipe the catheter with acetone-based solutions, tincture of iodine or polyeth-

ylene glycol-containing ointments. These can damage the polyurethane material if used 
over time.

• 

(Pediatric) Insertion techniques and placement locations are often modified according 

to the size and developmental age of the child. Only clinicians experienced in proper 
positioning and placement of midlines in pediatric patients should place this catheter in 
this patient population.

Содержание PowerMidline

Страница 1: ...1 Instructions For Use BARD...

Страница 2: ...or past treatment will prevent proper device stabilization and or access Warnings General Warnings When using alcohol or alcohol containing antiseptics with polyurethane midlines care should be taken...

Страница 3: ...dure for a particular patient A suitable trained clinician is responsible for evaluat ing the health status of a patient as it pertains to a power injection procedure Failure to warm contrast media to...

Страница 4: ...the dose Do not infuse against resistance To reduce potential for blood backflow into the catheter tip always remove syringes slowly while injecting the last 0 5 mL of sterile saline Do not reinsert n...

Страница 5: ...round any part of the catheter s tubing shaft or extension legs If using sutures to secure catheter USE THE SUTURE WINGS and make sure they do not occlude puncture or cut the catheter Do not use sciss...

Страница 6: ...Erosion Through the Skin Catheter Embolism Catheter Occlusion Catheter Related Sepsis Exit Site Infection Exit Site Necrosis Extravasations Fibrin Sheath Formation Hematoma Heparin Induced Thrombocyto...

Страница 7: ...lease tourniquet D Assess the selected vein to ensure the vessel size is adequate to accommodate the catheter being placed 2 Patient Position Catheter Measurement A Position the arm at a 90 angle B Fo...

Страница 8: ...tylet assembly as one unit Caution Catheter stylet must be wetted prior to stylet repositioning or withdrawal D Retract the stylet to well behind the point the catheter is to be cut Caution Do not all...

Страница 9: ...nit to prevent the needle from damaging or shearing the guidewire F Advance the small sheath and dilator together as a unit over the guidewire using a slight rotational motion If necessary a small inc...

Страница 10: ...Stabilize the catheter position by applying pressure to the vein distal to the introducer sheath B Withdraw the introducer sheath from the vein and away from the site C Split the introducer sheath an...

Страница 11: ...er to Instructions For Use on the proper use and removal The StatLock Stabilization Device should be monitored daily and replaced at least every seven days Caution The catheter must be secured in plac...

Страница 12: ...be tween hub and wings Chevron anchor tape on top of transparent dressing Tape Strip Securement Procedure Single Lumen 1 Place 1st anchor tape over wings or bifurcation 2 Cover site and 1st anchor tap...

Страница 13: ...emperature prior to power injec tion may result in catheter failure Warning Use of lumens not marked Power Injectable for power injection may result in catheter failure G Complete power injection stud...

Страница 14: ...LLER THAN 10 mL TO FLUSH AND CONFIRM PATENCY Patency should be assessed with a 10 mL or larger syringe with sterile saline Upon confirmation of patency administration of medication should be given in...

Страница 15: ...nd Toxic Reactions Adverse Events During Infusion Therapy Symposium University of California 1993 2 Findlay Steven R et al Hyperosmolar Triggering of Histamine Release from Human Basophils Journal of...

Страница 16: ...daccess com Clinical Information 1 800 443 3385 USA Email medical services crbard com STERILIZE Manufacturer Do Not Re Use Non Pyrogenic Reorder Number Lot Number Do Not Re sterilize Length Use By Ste...

Отзывы: