D.  If resistance is felt, stop removal. Apply warm compress and wait 20-30 minutes.

E.  Resume removal procedure.
F.  Inspect the end of the catheter and the catheter length to ensure complete removal.

Caution: 

After use, this product may be a potential biohazard. Handle and dispose 

of in accordance with medical practice and applicable local, state and federal laws 

and regulations.

WARNING:

 Intended for Single Use. DO NOT REUSE. Reuse and/or repackaging 

may create a risk of patient or user infection, compromise the structural integrity 

and/or essential material and design characteristics of the device, which may lead to 
device failure, and/or lead to injury, illness or death of the patient.

Caution:

 Do not use scissors to remove dressing to minimize the risk of cutting 

catheter.

References

      1.  Halpern M.D., Georges. “Allergic and Toxic Reactions.” Adverse Events    

              

     During Infusion Therapy Symposium, University of California, (1993)

      2.  Findlay, Steven R. et al,. “Hyperosmolar Triggering of Histamine Release from         

     Human Basophils.” Journal of Clinical Investigation. (1981)

      3.  Benjamin, Richard J. et al,. “Skin disinfection with single-step 2% chlorhexidine  

 

     swab is more effective than a two-step povidoneiodine method in preventing  

 

     bacterial contamination of apheresis platelets.” Transfusion. (2010)

      4.  “FDA Public Health Notice: Potential Hypersensitivity to chlorhexidine-Impregnated          

     Medical Devices.” FDA U.S. Food and Drug Administration. <http://www.fda.gov>.      
     (accessed March, 1998).

      5.  Beaudouin, E. et al,. “Immediate Hypersensitivity to chlorhexidine: literature review.”      

     European Annals of Allergy and Clinical Immunology. 36, no. 4 (2004)

      6.  Gorski, Lisa. et al,. “Infusion Therapy Standards of Practice.” 

     Journal of Infusion Nursing. (2016)

Revised date: November 2016

Bard, PowerMidline, and StatLock are trademarks and/or registered trademarks of C. R. 
Bard, Inc. 

© 2016 C. R. Bard, Inc. All rights reserved.

Содержание PowerMidline

Страница 1: ...1 Instructions For Use BARD...

Страница 2: ...or past treatment will prevent proper device stabilization and or access Warnings General Warnings When using alcohol or alcohol containing antiseptics with polyurethane midlines care should be taken...

Страница 3: ...dure for a particular patient A suitable trained clinician is responsible for evaluat ing the health status of a patient as it pertains to a power injection procedure Failure to warm contrast media to...

Страница 4: ...the dose Do not infuse against resistance To reduce potential for blood backflow into the catheter tip always remove syringes slowly while injecting the last 0 5 mL of sterile saline Do not reinsert n...

Страница 5: ...round any part of the catheter s tubing shaft or extension legs If using sutures to secure catheter USE THE SUTURE WINGS and make sure they do not occlude puncture or cut the catheter Do not use sciss...

Страница 6: ...Erosion Through the Skin Catheter Embolism Catheter Occlusion Catheter Related Sepsis Exit Site Infection Exit Site Necrosis Extravasations Fibrin Sheath Formation Hematoma Heparin Induced Thrombocyto...

Страница 7: ...lease tourniquet D Assess the selected vein to ensure the vessel size is adequate to accommodate the catheter being placed 2 Patient Position Catheter Measurement A Position the arm at a 90 angle B Fo...

Страница 8: ...tylet assembly as one unit Caution Catheter stylet must be wetted prior to stylet repositioning or withdrawal D Retract the stylet to well behind the point the catheter is to be cut Caution Do not all...

Страница 9: ...nit to prevent the needle from damaging or shearing the guidewire F Advance the small sheath and dilator together as a unit over the guidewire using a slight rotational motion If necessary a small inc...

Страница 10: ...Stabilize the catheter position by applying pressure to the vein distal to the introducer sheath B Withdraw the introducer sheath from the vein and away from the site C Split the introducer sheath an...

Страница 11: ...er to Instructions For Use on the proper use and removal The StatLock Stabilization Device should be monitored daily and replaced at least every seven days Caution The catheter must be secured in plac...

Страница 12: ...be tween hub and wings Chevron anchor tape on top of transparent dressing Tape Strip Securement Procedure Single Lumen 1 Place 1st anchor tape over wings or bifurcation 2 Cover site and 1st anchor tap...

Страница 13: ...emperature prior to power injec tion may result in catheter failure Warning Use of lumens not marked Power Injectable for power injection may result in catheter failure G Complete power injection stud...

Страница 14: ...LLER THAN 10 mL TO FLUSH AND CONFIRM PATENCY Patency should be assessed with a 10 mL or larger syringe with sterile saline Upon confirmation of patency administration of medication should be given in...

Страница 15: ...nd Toxic Reactions Adverse Events During Infusion Therapy Symposium University of California 1993 2 Findlay Steven R et al Hyperosmolar Triggering of Histamine Release from Human Basophils Journal of...

Страница 16: ...daccess com Clinical Information 1 800 443 3385 USA Email medical services crbard com STERILIZE Manufacturer Do Not Re Use Non Pyrogenic Reorder Number Lot Number Do Not Re sterilize Length Use By Ste...

Отзывы: