Flow Performance Guidelines For Bard Access Systems Power-Trialysis

  

Short-Term Dialysis Catheter

Performance Guidelines: (As suggested by In Vitro data)

:

Power Injection Flow Information (Injection pressure should be set at a maximum of 300 psi) 

Center Lumen

Configuration

Max Indicated

Power Injections Flow-Rate

1

Average Max Catheter

Pressure During Max Indicated Power 

Injection Rate

2

30 cm Straight & Curved Extension Legs

5 mL/sec

107 psi

24 cm Alphacurve

5 mL/sec

126 psi

1

 Represents maximum indicated flow rate for power injection of contrast media.

Internal catheter pressure during power injection of contrast media at 11.8cP with safety cut-off at 300 psi.

Maximum Static Burst Pressure Information

3

Center Lumen

Configuration

Average Max Static Burst Pressure

Range of Max Static Burst Pressures

30 cm Straight & Curved Extension Legs

358 psi

356-360 psi

24 cm Alphacurve

333 psi

315-348 psi

3

 Maximum burst pressure is the static burst pressure failure point of the power injectable lumen of the catheter when fully occluded. The 

margin of safety for power injection is demonstrated comparing the average static burst pressure to the catheter pressure during power 

injection at 5 mL/sec flow rate, assuming patency of the catheter has been verifed.

Dialysis Lumen Flow Information (Flow Rate vs. Lumen Pressure ) 

5

Average Pressures

30 cm Straight 

(400 mL/min Flow Rate)

24 cm Curved Ext. Legs 

(400 mL/min Flow Rate)

24 cm Alphacurve

 

(350 mL/min Flow Rate)

4

Venous (mmHg)

216

197

239

Arterial (mmHg)

-221

-204

-216

 

In Vitro flow data indicates flow rates above 350 mL/min for the Alphacurve configuration may exceed a prepump pressure of - 250 mmHg 

as per K/DOQI (2006) guidelines.

 

Flow Rate vs Lumen Pressure data obtained using a blood simulant of water and glycerin at 3.4cP.

Center Lumen Flow Information (Gravity Flow Rate (1 Meter of Water Infusion Pressure))

Configuration

Insertion Length

30 cm

24 cm

20 cm

15 cm

12.5 cm

Straight & Curved Extension Legs (mL/hr)

2258

2634

2830

3333

3721

Alphacurve

 (mL/hr)

2149

2387

2539

3092

Priming Volumes

Configuration

Insertion Length

30 cm

24 cm

20 cm

15 cm

12.5 cm

Straight & Curved Extension Legs (mL)

1.8

1.6

1.4

1.2

1.1

Alphacurve

 (mL)

2.0

1.8

1.6

1.5

Recirculation Rates

6

Configuration

Forward

Reverse

Straight & Curved Extension Legs

< 2%

< 2%

Alphacurve

< 2%

< 2%

 

Recirculation Rates obtained using a saline dilution technique with a blood simulant of water and glycerin at 3.4 cP.

Содержание Power-Trialysis

Страница 1: ...or Use SHORT TERM DIALYSIS CATHETERS SHORT TERM DIALYSIS CATHETERS Bard Access Systems Inc 605 North 5600 West Salt Lake City UT 84116 USA 1 801 522 5000 Clinical Information Hotline 1 800 443 3385 Or...

Страница 2: ...blood return and vigorously flush the catheter with the full 10 mL of sterile normal saline WARNING To prevent systemic heparanization of the patient the heparin solution must be aspirated out of the...

Страница 3: ...e grades of Pinch off that should be recognized with appropriate chest x ray as follows 2 The catheter must not be left in the femoral vein longer than three days It is recommended that jugular and su...

Страница 4: ...r Injectable for power injection of contrast media Power injection machine pressure limiting feature may not prevent over pressurization of an occluded catheter which may lead to cath eter failure Exc...

Страница 5: ...to expose the side of the groin to be accessed 2 Prepare the access site using standard surgical technique and drape the prepped area with sterile towels If hair removal is necessary use clippers or d...

Страница 6: ...e 2 Secure the wing onto the catheter by tying sutures around the wing using the suture grooves 3 Secure the removable wing in place by suturing through the holes or by using adhesive wound closures W...

Страница 7: ...lution injection is necessary for 48 72 hours provided the catheter has not been aspirated or flushed 3 Maintain patency of distal purple lumen per institution protocol for central lines 4 Clean cathe...

Страница 8: ...the average static burst pressure to the catheter pressure during power injection at 5 mL sec flow rate assuming patency of the catheter has been verifed Dialysis Lumen Flow Information Flow Rate vs...

Страница 9: ...ity of Bard Access Systems Inc under this limited product warranty does not extend to any abuse or misuse of this product or its repair by anyone other than an authorized Bard Access Systems Inc repre...