Bard Power-Trialysis Скачать руководство пользователя страница 4

Warnings 

(cont.)

•  To avoid damage to vessels and viscus, prolonged infusion pressures must not exceed 25 psi (172 kPa).  The use of a 10 mL or larger 

syringe is recommended because smaller syringes generate more pressure than larger syringes.

 

NOTE:  A three pound (13.3 Newton) force on the plunger of a 3 mL syringe generates pressure in excess of 30 psi (206 kPa) whereas the 

same three pound (13.3 Newton) force on the plunger of a 10 mL syringe generates less than 15 psi (103 kPa) of pressure.

•  Accessories and components used in conjunction with this catheter must incorporate Luer-lock adapters in order to avoid inadvertent 

disconnection.

•  Catheters should be implanted carefully to avoid any sharp or acute angles which could compromise the opening of the catheter lumens.

•  Before dialysis begins, all connections to the extracorporeal circuit must be checked carefully.  During all dialysis procedures frequent 

visual inspection must be conducted to detect leaks and prevent blood loss or entry of air into the extracorporeal circuit.  Excess blood 

leakage may lead to patient shock.

•  In the rare event of a leak, the catheter must be clamped immediately.  Necessary remedial action must be taken prior to resuming dialysis 

procedure.

•  If the catheter is not used immediately for treatment, follow the suggested Catheter Patency Guidelines.

•  Failure to clamp extensions when not in use may lead to air embolism.

•  Verification of the catheter tip location must be confirmed by x-ray to ensure proper placement.

•  To prevent systemic heparinization of the patient, the heparin solution must be aspirated out of the lumens immediately prior to using 

the catheter.

•  For jugular and subclavian insertion, the patient must be placed on a cardiac monitor during this procedure.  Cardiac arrhythmia may 

result if the guidewire is allowed to pass into the right atrium.  The guidewire must be held securely during the procedure.

•  The risk of infection is increased with femoral vein insertion.

•  Recirculation in femoral catheters was reportedly significantly greater than in internal jugular catheters. 

4

•  Before attempting the insertion of the catheter, ensure that you are familiar with the possible complications listed below and their emer-

gency treatment should they occur.

•  Cannulation of the left internal jugular vein was reportedly associated with a higher incidence of complications compared to catheter 

placement in the right internal jugular vein. 

3

 

•  Failure to warm contrast media to body temperature (37°C) prior to power injection may result in catheter failure 

•  Failure to ensure patency of the catheter prior to power injection studies may result in catheter failure.

•  Do not use lumens not marked “Power Injectable” for power injection of contrast media.

•  Power injection machine pressure limiting feature may not prevent over-pressurization of an occluded catheter, which may lead to cath-

eter failure.

•  Exceeding the maximum flow rate of 5 mL/sec or the maximum pressure of power injectors of 300 psi may result in catheter failure and/

or catheter tip displacement.

•  Catheter indication for power injection of contrast media implies the catheter’s ability to withstand the procedure, but does not imply 

appropriateness of the procedure for a particular patient.  A suitably trained clinician is responsible for evaluating the health status of a 

patient as it pertains to a power injection procedure.

•  Do not power inject through a catheter that exhibits signs of clavicle-first rib compression or pinch-off, as it may result in catheter failure.

•  If local pain, swelling, or signs of extravasation are noted, the injection should be stopped immediately.

•  After use, this product may be a potential biohazard.  Handle and discard in accordance with accepted medical practice and applicable 

local, state and federal laws and regulations.

•  CVP Monitoring should always be used in conjunction with other patient assessment metrics when evaluating cardiac function.

•  CVP Monitoring should not be performed during hemodialysis, hemoperfusion or apheresis.

•  For optimal product performance and to avoid complications, do not insert any portion of the curve into the vein.

Cautions

•  Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.

•  Carefully read and follow all instructions prior to use.

•  Only qualified health care practitioners should insert, manipulate, and remove these devices.

•  Strict aseptic technique must be used during the insertion, maintenance, and catheter removal procedures.

•  Do not pull back guidewire over needle bevel as this may sever the end of the guidewire.  The introducer needle must be removed first.  

Also, if unusual resistance is met during manipulation of the guidewire, discontinue the procedure and determine the cause of resis-

tance before proceeding.  Withdraw needle and guidewire if cause of resistance cannot be determined.

•  Do not allow the guidewire to inadvertently advance totally into the vessel.

•  For jugular and subclavian insertion, the catheter tip should not be located in the right atrium.

•  Sterile, non-pyrogenic unless package is damaged or opened.

•  Left sided placement in particular, may provide unique challenges due to the right angles formed by the innominate vein and at the 

left brachiocephalic junction with the left SVC.

3

•  Follow all contraindications, warnings, precautions, and instructions for all infusates, including contrast media, as specified by their 

manufacturer.

•  In vitro flow data indicates flow rates above 350 mL/min for Alphacurve™ Catheter configurations may exceed -250 mmHg prepump 

pressure as per K/DOQ1 (2006) guidelines.

Содержание Power-Trialysis

Страница 1: ...or Use SHORT TERM DIALYSIS CATHETERS SHORT TERM DIALYSIS CATHETERS Bard Access Systems Inc 605 North 5600 West Salt Lake City UT 84116 USA 1 801 522 5000 Clinical Information Hotline 1 800 443 3385 Or...

Страница 2: ...blood return and vigorously flush the catheter with the full 10 mL of sterile normal saline WARNING To prevent systemic heparanization of the patient the heparin solution must be aspirated out of the...

Страница 3: ...e grades of Pinch off that should be recognized with appropriate chest x ray as follows 2 The catheter must not be left in the femoral vein longer than three days It is recommended that jugular and su...

Страница 4: ...r Injectable for power injection of contrast media Power injection machine pressure limiting feature may not prevent over pressurization of an occluded catheter which may lead to cath eter failure Exc...

Страница 5: ...to expose the side of the groin to be accessed 2 Prepare the access site using standard surgical technique and drape the prepped area with sterile towels If hair removal is necessary use clippers or d...

Страница 6: ...e 2 Secure the wing onto the catheter by tying sutures around the wing using the suture grooves 3 Secure the removable wing in place by suturing through the holes or by using adhesive wound closures W...

Страница 7: ...lution injection is necessary for 48 72 hours provided the catheter has not been aspirated or flushed 3 Maintain patency of distal purple lumen per institution protocol for central lines 4 Clean cathe...

Страница 8: ...the average static burst pressure to the catheter pressure during power injection at 5 mL sec flow rate assuming patency of the catheter has been verifed Dialysis Lumen Flow Information Flow Rate vs...

Страница 9: ...ity of Bard Access Systems Inc under this limited product warranty does not extend to any abuse or misuse of this product or its repair by anyone other than an authorized Bard Access Systems Inc repre...

Страница 10: ......

Отзывы: