Nio

Supported displays:

E-3620, MDNG-5121, MDNG-2121,

MDNC-3121, MDNG-6121, MDNC-2121

Содержание Nio E-3620

Страница 1: ...User Guide Nio Supported displays E 3620 MDNG 5121 MDNG 2121 MDNC 3121 MDNG 6121 MDNC 2121 ...

Страница 2: ... This page intentionally left blank 2 Nio ...

Страница 3: ...isplay system improves your diagnostic reading routine for a multitude of medical imaging applications Following symbols may be used throughout this guide Warning Risk of injury to human beings Caution Risk of damage to the product Important notice or remark Note Hint tip Additional information Read all the safety information starting on page 43 before operating your BARCO NIO Display WARNING ...

Страница 4: ... This page intentionally left blank 4 Nio ...

Страница 5: ... VESA mount installation 18 Daily operation 20 Recommendations for daily operation 20 Stand by switching 22 Locking and unlocking user controls 23 Bringing up the OSD menu 25 Navigating through the OSD menus 26 Making selections in the OSD menus 26 Changing values in the OSD menus 27 Advanced operation 28 On screen display OSD 28 Input Selection menu 28 Luminance and color menu 29 Display Function...

Страница 6: ... 37 Switching DPMS on off 40 Cleaning your display 41 Front glass 41 Cabinet 42 Important information 43 Safety information 43 Environmental information 46 Regulatory compliance information 49 Explanation of symbols 51 Legal disclaimer 52 Technical specifications 53 ...

Страница 7: ...le MDNC 2121 only a USB cable a set of AC power cords an external power supply If you ordered a Barco display controller it s also in the box together with its accessories A dedicated user guide is available on the system CD Keep your original packaging It is designed for this display and is the ideal protection during transport ...

Страница 8: ...8 Nio Parts controls and connectors Front view Figure 1 Front side Power LED USB downstream port Control wheel BARCO 1 2 3 1 2 3 ...

Страница 9: ... Figure 2 Rear side Connector compartment cover Tilt swivel foot cover Tilt swivel foot Slot for security cable DVI digital video input USB upstream port USB downstream port DC power input Tilt swivel foot clip 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 ...

Страница 10: ...igure 3 Rear side Connector compartment cover Tilt swivel foot cover Tilt swivel foot DC power input 5 Vdc Slot for security cable DVI digital video input USB upstream port USB downstream port Tilt swivel foot clip 1 2 3 4 5 7 6 8 9 10 1 2 3 4 5 6 7 8 9 10 ...

Страница 11: ...nector compartment cover Tilt swivel foot cover Tilt swivel foot Slot for security cable DisplayPort video input DVI digital video input USB upstream port USB downstream port DC power input Tilt swivel foot clip 1 2 3 4 5 7 6 8 9 10 1 2 3 4 5 6 7 8 9 10 ...

Страница 12: ...ip TO REMOVE THE CLIP 1 Position the display with its rear side facing you 2 Pull the red clip out of the fixation holes in the foot 3 Keep the clip in case the display needs to be shipped later Prior to installing your BARCO NIO Display and connecting all necessary cables make sure to have a suitable display controller physically installed in your computer If you are using a Barco display control...

Страница 13: ...ition You can change the orientation of the panel at any time but it is more convenient to select landscape or portrait orientation before connecting the cables Figure 6 Portrait orientation Figure 7 Landscape orientation BARCO BARCO ...

Страница 14: ...f the panel and take the panel at both sides 2 Very important Til the panel before changing the orientation Figure 8 Tilt the panel before rotating Should you change the panel orientation without tilting it first you might irreversibly damage the tilt swivel mechanism ...

Страница 15: ...y or the system you change the panel orientation while an image is on the screen the result depends on the graphic board and the resolution of the image In some cases the image will be rotated automatically in other cases it will not be rotated e g when pixels would be lost after rotation If necessary change the image resolution in the display control panel and restart the system after changing th...

Страница 16: ... DisplayPort cable to the DVI DisplayPort input of the display 1 Connect the other end of the DVI DisplayPort cable to the DVI DisplayPort connector of the display controller board Connect a PC USB downstream connector to the display s USB upstream connector by means of a USB cable 2 Connect any USB device to one of the display s USB downstream connectors 3 Connect the DC power input 4 of the disp...

Страница 17: ...ear of the tilt swivel foot Bind the cables together above and under the foot by means of the 2 velcro strips attached to the inside of the foot cover packed inside the accessory box At last put the foot cover back in place Re attaching the covers Figure 13 Installing the cover 1 The foot cover is packaged separately in the accessory box Unpack the foot cover 2 Push the upper side of the cover ont...

Страница 18: ...the panel TO ATTACH THE DISPLAY TO AN ARM STAND 1 Put the display face down on a clean and soft surface Be careful not to damage the panel screen 2 Remove the tilt swivel foot cover Use an arm that is approved by VESA according to the VESA 100 mm standard Use an arm that can support a weight of at least 13 kg 28 66 lbs Never move a display attached to an arm by pulling or pushing the display itsel...

Страница 19: ... 14 Display with tilt swivel foot cover removed 4 Unscrew the 2 screws fixing the round plastic cover B 5 Lift up the round plastic cover 6 Remove the four screws fixing the foot while supporting the foot 7 Attach the arm stand firmly to the panel using 4 screws M4 x 8 mm Figure 15 Location of the screws A B 4 screws M4 x 8mm ...

Страница 20: ...d or significantly reduce the occurrence of this phenomenon by using a screen saver You can activate a screen saver in the Display properties window of your workstation In case you are working with the same image or an application with static image elements for several hours continuously so that the screen saver is not activated change the image content regularly to avoid image retention of the st...

Страница 21: ...hnology and missing pixels consult the dedicated white papers available at www barco com healthcare Barco recommends keeping color matched displays together Furthermore it is important to use all displays of a multi head configuration at the same rate to preserve color matching throughout the economic lifetime of the system Barco recommends to install MediCal QAWeb Agent and apply the default QAWe...

Страница 22: ... by The LED turns orange Figure 16 Location of the control wheel When the display is in stand by press the control wheel to switch it back on The connected power supply also provides a switch that can be used to turn the power completely off To use the display please make sure to switch on this power supply This can be done by pushing the on off switch on the power supply into the position BARCO P...

Страница 23: ...confirm and switch to stand by 4 The LED turns orange Figure 17 Location of the control wheel 5 When the display is in stand by press the control wheel shortly to switch it back on Locking and unlocking user controls The User Controls function allows to disable or enable the control wheel functions The procedure is slightly different depending on the type of display When user controls are disabled...

Страница 24: ...USER CONTROLS ON MDNC 2121 MDNC 3121 MDNG 2121 AND MDNG 6121 1 Rotate the control wheel to display the on screen display The Main Menu appears 2 Rotate the control wheel to select Settings 3 Press the control wheel to enter the Settings menu Figure 19 Settings menu example 4 Rotate the control wheel to select User Controls 5 Press the control wheel to activate the User Controls function This is in...

Страница 25: ...ep clockwise 5 Press the control wheel 2 times 6 Rotate the control wheel 1 step counterclockwise In MDNC 2121 MDNC 3121 MDNG 2121 and MDNG 6121 displays the user controls are automatically enabled after this procedure In other Nio displays you must enable the user controls again in the Settings menu Bringing up the OSD menu When the OSD is not on the screen rotate the control wheel to display the...

Страница 26: ...s to the previous higher level menu When you press on Exit in the main menu the OSD is hidden Figure 21 OSD menu navigation Making selections in the OSD menus 1 Navigate through the menus and highlight the parameter you wish to select 2 Press the control wheel to make a selection or to toggle between several settings BARCO NIO Display MAIN MENU Luminance Settings Information EXIT SETTINGS DPMS On ...

Страница 27: ...D menus 1 Navigate through the menus and highlight the parameter you wish to adjust 2 Press the control wheel to go into edit mode 3 Rotate the wheel to change the value 4 Press the wheel to enter the new value and return to the menu ...

Страница 28: ...ts the incoming video signal automatically and displays it on the screen DVI Always displays the DVI signal on the sreen When no DVI signal available there will be no image on the screen DP Always displays the DisplayPort signal on the screen When no DP signal available there will be no image on the screen ...

Страница 29: ...he luminance target See note below Color target Allows to select from a list of factory defined and user defined color targets You can select from a range of standard Daylight D color temperatures Clearbase Bluebase and Native The higher the D number the more bluish the color temperature is Native is the uncorrected color temperature of the LCD panel Viewing mode Allows to select a pre defined vie...

Страница 30: ...rget cannot be reached e g due to aging of the backlight the Measured luminance line changes to Minimum value reached or Maximum value reached Display function Allows to select from a list of pre defined display functions If the DICOM DF is selected an additional function is available ALC DICOM options Jumps to the ALC DICOM Options submenu which allows to edit the settings for the DICOM display f...

Страница 31: ...m the ambient light is not taken into account When set to Preset a preset ambient light value determined by the selected reading room is taken into account Reading room Allows to select from a pre defined list of reading room types You must select a reading room that corresponds to the type of room the display is installed in Reading room def Jumps to the reading room definition submenu which allo...

Страница 32: ...ue entered here you should take measures to darken the room or select another reading room type Preset corr value This value is taken into account in the calculation of the DICOM DF if DICOM Offset is set to Preset To each reading room type corresponds another preset value Preset Ambient value Shows the ambient light correction value taken into account to calculate the DICOM DF expressed in lux Br...

Страница 33: ... the user cannot display the OSD unit the user controls are enabled again For more information Locking and unlocking user controls on page 23 Language Allows to select another language for the OSD menus Auto menu exit Allows to switch the automatic menu exit on off When switched on the OSD is closed automatically when left idle for a certain time Preferred Input Select the video resolution you wis...

Страница 34: ...r is left idle for a certain time Barco recommends to switch DPMS on and to use a good Windows screen saver to prevent image burn in image retention on the LCD panel Product The display type Serial No Indicates the display serial number SW Version Displays the current internal software version Display Runtime Indicates the total time the display has been operating including the time in stand by Ba...

Страница 35: ...here CT scans are observed the CT MR NM reading room is the preferred selection The reading room is defined by 2 parameters Maximum Ambient Light defines the maximum light allowed in this type of room This value can be adjusted within certain limits determined by the selected reading room Preset Correction Value is the predefined correction value for this reading room This value can be adjusted wi...

Страница 36: ...You can select the following DFs DICOM Select a DICOM display function for most medical viewing applications The DICOM function results in more visible grayscales in the images Dynamic Gamma 2 2 Dynamic Gamma 1 8 These are gamma functions that are shifted to take into account the non zero luminance of an LCD panel when driven with a black signal They are especially useful in CT applications to imp...

Страница 37: ... Select the DICOM display function See To select a display function DF on page 36 2 Enter the ALC DICOM submenu 3 In the ALC DICOM Options submenu rotate the control wheel to select DICOM Offset Press the control wheel shortly to go into edit mode 4 Rotate the control wheel to select an offset 5 Press the control wheel shortly to activate the selection User This is a user definable DF Use MediCal ...

Страница 38: ...g room that corresponds to the room type the display is installed in Return to the menu 5 You can select the following reading room types Dark room The DICOM DF will be recalculated taking an ambient light value of 0 lux into account This means that the measured ambient light does not influence the DICOM DF Preset The DICOM DF will be recalculated taking a preset ambient light value into account T...

Страница 39: ...o rotate the control wheel to select Reading Room Press the control wheel shortly to go into edit mode Rotate the control wheel to select the reading room Press the control wheel shortly to confirm the selection CT MR NM Corresponds to light conditions in diagnostic reading rooms for computed tomography magnetic resonance or nuclear medicine scans Office Corresponds to light conditions in office r...

Страница 40: ... the control wheel to change the value Press the control wheel shortly to savethe changes You can change this value within certain limits determined by the selected reading room Switching DPMS on off TO SWITCH DPMS ON OFF 1 When there is no OSD on the screen rotate the control wheel The main OSD menu appears 2 Rotate the control wheel to select the Settings menu Press the control wheel shortly to ...

Страница 41: ...water Flux Pril Windex Glass Plus Ajax glass cleaner Bohle glass cleaner Sidolin glass cleaner Mr Proper 25 ethanol solution EtOH Take care not to damage or scratch the front glass Be careful with rings or other jewelry and do not apply excessive pressure on the front glass Do not apply or spray liquid directly to the front glass as excess liquid may cause damage to internal electronics Instead ap...

Страница 42: ...e cleaning products is available use plain water 2 Repeat with water only 3 Wipe dry with a dry cloth Do not use following products Alcohol solvents at higher concentration 5 Strong alkalis lye strong solvents Acid Detergents with fluoride Detergents with ammonia Detergents with abrasives Steel wool Sponge with abrasives Steel blades Cloth with steel thread Do not apply or spray liquid directly to...

Страница 43: ...ctions manual Follow all instructions for operation and use ELECTRICAL SHOCK TYPE OF PROTECTION ELECTRICAL Equipment with external power supply Class I equipment DEGREE OF SAFETY FLAMMABLE ANESTHETIC MIXTURE Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide NON PATIENT CARE EQUIPMENT Equipment primarily for use in a health car...

Страница 44: ... OVER VOLTAGE If the equipment is not used for a long time disconnect it from the AC inlet to avoid damage by transient over voltage To fully disengage the power to the unit please disconnect the power cord from the AC inlet POWER CORDS Utilize a UL listed detachable power cord 3 wire type SJ or equivalent 18 AWG min rated 300 V min provided with a hospital grade type plug 5 15P configuration for ...

Страница 45: ...ou use an unstable cart or stand the equipment may fall causing serious injury to a child or adult and serious damage to the equipment THIS APPARATUS CONFORMS TO CE0120 MDD 93 42 EEC class IIb product IEC 60601 1 UL 60601 1 CAN CSA C22 2 No 601 01 M90 c UL CCC GB4943 1995 IEC 60950 1 IEC 60601 1 2005 EN 60601 1 2 2006 ANSI AAMI ES 60601 1 2005 CAN CSA C22 2 No 60601 1 08 MDNC 2121 Medical equipmen...

Страница 46: ...om uncontrolled waste disposal please separate these items from other types of waste and recycle them responsibly to promote the sustainable reuse of material resources For more information about recycling of this product please contact your local city office your municipal waste disposal service or the shop where you purchased the product MERCURY NOTICE This Barco product consists of materials th...

Страница 47: ...Barco s product may contain The RoHS of Chinese Mainland is included in the MCV standard of the Ministry of Information Industry of China in the section Limit Requirements of toxic substances in Electronic Information Products 零件项目 名称 Component Name 有毒有害物质或元素 Hazardous Substances or Elements 铅 Pb 汞 Hg 镉 Cd 六价铬 Cr6 多溴联苯 PBB 多溴二苯醚 PBDE 印制电路配件 Printed Circuit Assemblies X O O O O O 液晶面板 LCD panel X X...

Страница 48: ...o is based on the Standard of Electronic Information Products Environmental Friendly Use Period of Chinese Mainland 电源供应器 Power supply unit X O O O O O 文件说明书 Paper manuals O O O O O O 光盘说明书 CD manual O O O O O O O 表示该有毒有害物质在该部件所有均质材料中的含量均在 SJ T 11363 2006 标准规定的限 量要求以下 O Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit r...

Страница 49: ... of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to prov...

Страница 50: ...ct to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Interference can be determined by turning the equipment off and on If this equipment does cause harmful interference to or suffer from harmful interference of surrounding equipment the user is encourage...

Страница 51: ...according to the UL regulations or Indicates the display is approved according to the c UL regulations Indicates the display is approved according to the DEMKO regulations Indicates the display is approved according to the CCC regulations Indicates the display is approved according to the VCCI regulations Indicates the USB connectors on the display Indicates the manufacturing date Indicates the te...

Страница 52: ...sclosure of a Barco software product in any form is authorized The specifications of Barco products are subject to change without notice TRADEMARKS All trademarks and registered trademarks are property of their respective owners COPYRIGHT NOTICE This document is copyrighted All rights are reserved Neither this document nor any part of it may be reproduced or copied in any form or by any means grap...

Страница 53: ... 4 3 Resolution 2MP 1600 x 1200 Pixel pitch 0 2700 mm Color imaging No Gray imaging Yes Number of grayscales LUT in LUT out 1024 grey levels 10 12 Viewing angle H V 170 Uniform Luminance Technology ULT Yes Per Pixel Uniformity PPU No Ambient Light Compensation ALC No Backlight Output Stabilization BLOS Yes I Guard No Maximum luminance 1650 cd m DICOM calibrated luminance ULT off 500 cd m Contrast ...

Страница 54: ...nt DVI DMPM Dot clock 280 MHz OSD languages English French German Spanish Italian Dimensions with stand W x H x D Portrait 385 x 525 585 x 250 mm Landscape 485 x 475 535 x 250 mm Dimensions w o stand W x H x D Portrait 385 x 485 x 115 mm Landscape 485 x 385 x 115 mm Dimensions packaged W x H x D 500 x 675 x 320 mm Net weight with stand 12 2 kg Net weight w o stand 8 14 kg Net weight packaged with ...

Страница 55: ... Class B KETI IEC 60601 1 2005 EN 60601 1 2 2006 ANSI AAMI ES 60601 1 2005 CAN CSA C22 2 No 60601 1 08 Supplied accessories User Guide Quick Installation Sheet Video cable DVI Dual Link Mains cables UK European CEBEC KEMA USA UL CSA adaptor plug NEMA 5 15P Chinese CCC USB 2 0 cable External power supply Optional accessories NA QA software MediCal QAWeb Unit per pallet NA Pallet dimensions NA Warra...

Страница 56: ...07 mm Color imaging No Gray imaging Yes Number of grayscales LUT in LUT out 1024 grey levels 10 10 Viewing angle H V 170 Uniform Luminance Technology ULT Yes Per Pixel Uniformity PPU No Ambient Light Compensation ALC No Backlight Output Stabilization BLOS Yes I Guard No Maximum luminance 1000 cd m DICOM calibrated luminance ULT off 500 cd m Contrast ratio typical 900 1 Response time Tr Tf 50 ms Sc...

Страница 57: ...OSD languages English Dimensions with stand W x H x D Portrait 385 x 525 585 x 250 mm Landscape 485 x 475 535 x 250 mm Dimensions w o stand W x H x D Portrait 385 x 485 x 115 mm Landscape 485 x 385 x 115 mm Dimensions packaged W x H x D 500 x 675 x 320 mm Net weight with stand 13 kg Net weight w o stand 9 kg Net weight packaged with stand 15 kg Net weight packaged w o stand 11 kg Height adjustment...

Страница 58: ...RoHs MDD WEEE CE 0120 VCCI FCC Class B KETI IEC 60601 1 2005 EN 60601 1 2 2006 ANSI AAMI ES 60601 1 2005 CAN CSA C22 2 No 60601 1 08 Supplied accessories User Guide Quick Installation Sheet Video cable DVI Dual Link Mains cables UK European CEBEC KEMA USA UL CSA adaptor plug NEMA 5 15P Chinese CCC External power supply Optional accessories NA QA software MediCal QAWeb Unit per pallet NA Pallet dim...

Страница 59: ...ch 0 165 mm Color imaging No Gray imaging Yes Number of grayscales LUT in LUT out 10 bit 10 bit Viewing angle H V 170 Uniform Luminance Technology ULT No Per Pixel Uniformity PPU No Ambient Light Compensation ALC No Backlight Output Stabilization BLOS Yes I Guard No Maximum luminance 700 cd m DICOM calibrated luminance ULT off 500 cd m Contrast ratio typical 800 1 Response time Tr Tf 50 ms Scannin...

Страница 60: ...Hz OSD languages English Dimensions with stand W x H x D Portrait 408 x 489 549 x 250 mm Landscape 492 x 531 591 x 250 mm Dimensions w o stand W x H x D Portrait 408 x 492 x 115 mm Landscape 492 x 408 x 115 mm Dimensions packaged W x H x D 500 x 320 x 675 mm Net weight with stand 12 5 kg Net weight w o stand 8 5 kg Net weight packaged with stand 15 2 kg Net weight packaged w o stand NA Height adju...

Страница 61: ...MDD WEEE CE 0120 VCCI FCC Class B KETI IEC 60601 1 2005 EN 60601 1 2 2006 ANSI AAMI ES 60601 1 2005 CAN CSA C22 2 No 60601 1 08 Supplied accessories User Guide Quick Installation Sheet Video cable DVI Dual Link Mains cables UK European CEBEC KEMA USA UL CSA adaptor plug NEMA 5 15P Chinese CCC USB 2 0 cable External power supply Optional accessories NA QA software MediCal QAWeb Unit per pallet NA P...

Страница 62: ...l pitch 0 2115 mm Color imaging Yes Gray imaging No Color support 24 bit Viewing angle H V 176 Uniform Luminance Technology ULT Yes Per Pixel Uniformity PPU No Ambient Light Compensation ALC No Backlight Output Stabilization BLOS Yes I Guard No Maximum luminance 800 cd m DICOM calibrated luminance ULT off 400 cd m Contrast ratio typical 750 1 Response time Tr Tf 24 ms Scanning frequency H V 15 128...

Страница 63: ...an Spanish Italian Dimensions with stand W x H x D Portrait 385 x 525 585 x 250 mm Landscape 485 x 475 535 x 250 mm Dimensions w o stand W x H x D Portrait 385 x 485 x 115 mm Landscape 485 x 385 x 115 mm Dimensions packaged W x H x D 500 x 320 x 675 mm Net weight with stand 13 kg Net weight w o stand 9 kg Net weight packaged with stand 15 kg Net weight packaged w o stand NA Height adjustment range...

Страница 64: ...I IEC 60601 1 2005 EN 60601 1 2 2006 ANSI AAMI ES 60601 1 2005 CAN CSA C22 2 No 60601 1 08 Supplied accessories User Guide Quick Installation Sheet Video cable DVI Dual Link Mains cables UK European CEBEC KEMA USA UL CSA adaptor plug NEMA 5 15P Chinese CCC USB 2 0 cable External power supply Optional accessories NA QA software MediCal QAWeb Unit per pallet NA Pallet dimensions NA Warranty 5 years ...

Страница 65: ...1540 mm Color imaging No Gray imaging Yes Number of grayscales LUT in LUT out 1024 grey levels 10 12 Viewing angle H V 170 Uniform Luminance Technology ULT Yes Per Pixel Uniformity PPU No Ambient Light Compensation ALC No Backlight Output Stabilization BLOS Yes I Guard No Maximum luminance 1100 cd m DICOM calibrated luminance ULT off 500 cd m Contrast ratio typical 1000 1 Response time Tr Tf 32 ms...

Страница 66: ...English French German Spanish Italian Dimensions with stand W x H x D Portrait 385 x 525 585 x 250 mm Landscape 485 x 475 535 x 250 mm Dimensions w o stand W x H x D Portrait 385 x 485 x 115 mm Landscape 485 x 385 x 115 mm Dimensions packaged W x H x D 500 x 320 x 675 mm Net weight with stand 12 3 kg Net weight w o stand 8 3 kg Net weight packaged with stand 17 3 kg Net weight packaged w o stand 1...

Страница 67: ... VCCI FCC Class B KETI IEC 60601 1 2005 EN 60601 1 2 2006 ANSI AAMI ES 60601 1 2005 CAN CSA C22 2 No 60601 1 08 Supplied accessories User Guide Quick Installation Sheet Video cable DVI Dual Link Mains cables UK European CEBEC KEMA USA UL CSA adaptor plug NEMA 5 15P Chinese CCC USB 2 0 cable External power supply Optional accessories NA QA software MediCal QAWeb Unit per pallet NA Pallet dimensions...

Страница 68: ...ch 0 2700 mm 0 01063 Color imaging Yes Gray imaging No Color support 24 bit Viewing angle H V 176 Uniform Luminance Technology ULT Yes Per Pixel Uniformity PPU No Ambient Light Compensation ALC No Backlight Output Stabilization BLOS Yes I Guard No Maximum luminance 750 cd m DICOM calibrated luminance ULT off 400 cd m Contrast ratio typical 1050 1 Response time Tr Tf 35 ms Scanning frequency H V 15...

Страница 69: ... Dot clock 165 MHz OSD languages English French German Spanish Italian Dimensions with stand W x H x D Portrait 385 x 525 585 x 250 mm Landscape 485 x 475 535 x 250 mm Dimensions w o stand W x H x D Portrait 385 x 485 x 115 mm Landscape 485 x 385 x 115 mm Dimensions packaged W x H x D 500 x 320 x 675 mm Net weight with stand 13 kg Net weight w o stand 9 kg Net weight packaged with stand 15 kg Net ...

Страница 70: ...k Installation Sheet Video cables 1 x DVI Single Link 1x DisplayPort Mains cables UK European CEBEC KEMA USA UL CSA adaptor plug NEMA 5 15P Chinese CCC USB 2 0 cable External power supply Optional accessories NA QA software MediCal QAWeb Unit per pallet NA Pallet dimensions NA Warranty 5 years Operating temperature 0 C to 40 C 15 C to 32 C within specs Storage temperature 20 C to 60 C Operating hu...

Страница 71: ...Nio 71 ...

Страница 72: ...0120 Barco n v President Kennedypark 35 8500 Kortrijk Belgium www barco com B410582 09 November 2012 ...