Nio
29
Luminance and color menu
* Only in MDNC-2121 and MDNC-3121 displays.
** Only in MDNC-2121, MDNC-3121, MDNG-2121 and MDNG-6121
displays
In grayscale Nio displays this menu is called “Luminance”.
Measured
luminance**
Indicates the actual luminance measured
by the internal sensor. This is a read-only
value. It is expressed in a percentage of the
calibrated value (100% is calibrated).
Luminance target
Allows to manually adjust the luminance
target. See note below.
Color target*
Allows to select from a list of factory-
defined and user-defined color targets.
You can select from a range of standard
Daylight (D) color temperatures, Clearbase,
Bluebase and Native.
The higher the D-number, the more bluish
the color temperature is.
‘Native’ is the uncorrected color
temperature of the LCD panel.
Viewing mode**
Allows to select a pre-defined viewing
mode.
Diagnostic
is the standard mode for
viewing medical images. In
Text
mode the
calibrated luminance is decreased so the
viewing conditions are optimal for viewing
office documents (text documents,
spreadsheets etc.)