The compliance of this monitor with Medical Safety and EMC requirements has been evaluated using the
external (optional) Skynet medical power supply model BAR-A159. If a different power supply will be used,
further investigation for Safety and EMC requirements, have to be performed at system level.
Power cords:
•
Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V min.,
provided with a hospital-grade type plug 5-15P configuration for 120V application, or 6-15P for 240V
application.
•
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
•
Mains lead protection (U.S.: Power cord): Power cords should be routed so that they are not likely to be
walked upon or pinched by items placed upon or against them, paying particular attention to cords at plugs
and receptacles.
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient over-
voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1 3rd
ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
Water and moisture
The equipment is IP21 compliant (IPx5 front side only) in vertical position.
The power supply is not approved for IP21. The power supply must be mounted in a flat position for
best resistance to fluids.
Moisture condensation
Do not use monitor under rapid temperature and humidity change condition or avoid cold air from air-
conditioning outlet directly.
Moisture may condense on the surface or inside of the unit, or create a mist residue inside the protection
plate, this is not a malfunction of the product itself, although it may cause damage to the monitor.
If condensation happens, let the monitor stand unplugged until there is no condensation.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a cupboard
or another enclosed location, heed the necessary space between the set and the sides of the cupboard.
Installation
•
Place the equipment on a flat, solid and stable surface that can support the weight of at least 3
equipments. If you use an unstable cart or stand, the equipment may fall, causing serious injury to a child
or adult, and serious damage to the equipment.
•
Do not allow to climb or rest on the equipment.
•
When adjusting the angle of the equipment, move it slowly so as to prevent the equipment from moving or
slipping off from its stand or arm.
•
When the equipment is attached to an arm, do not use the equipment as a handle or grip in order to move
the equipment. Please refer to the instruction manual of the arm for instructions on how to move the arm
with the equipment.
•
Provide full attention to safety during installation, periodic maintenance and examination of this equipment.
•
Sufficient expertise is required for installing this equipment, especially to determine the strength of the wall
for withstanding the display's weight. Be sure to entrust the attachment of this equipment to the wall to
licensed contractors of Barco and pay adequate attention to safety during the installation and usage.
Содержание MDSC-2324 DDI
Страница 1: ...ENABLING BRIGHT OUTCOMES MDSC 2324 LED MDSC 2324 DDI MDSC 2324 MNA User Guide MDSC 2324 ...
Страница 6: ...K5903252 451920612633 02 MDSC 2324 6 ...
Страница 7: ...7 K5903252 451920612633 02 MDSC 2324 Welcome 1 ...
Страница 10: ...K5903252 451920612633 02 MDSC 2324 10 Welcome ...
Страница 11: ...11 K5903252 451920612633 02 MDSC 2324 Product overview 2 ...
Страница 20: ...K5903252 451920612633 02 MDSC 2324 20 Product overview ...
Страница 21: ...21 K5903252 451920612633 02 MDSC 2324 Display installation 3 ...
Страница 30: ...K5903252 451920612633 02 MDSC 2324 30 Display installation ...
Страница 31: ...31 K5903252 451920612633 02 MDSC 2324 Daily operation 4 ...
Страница 41: ...41 K5903252 451920612633 02 MDSC 2324 Advanced operation 5 ...
Страница 59: ...59 K5903252 451920612633 02 MDSC 2324 Important information 6 ...