Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Coronis Fusion 4MP / 6MP is intended for use in the electromagnetic environment specified below. The
customer or the user of the Coronis Fusion 4MP / 6MP should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Coronis Fusion 4MP /
6MP requires continued
operation during power
mains interruptions, it is
recommended that the
Coronis Fusion 4MP /
6MP be powered from an
uninterruptible power
supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
5
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
5:
Coronis Fusion 4MP / 6MP doesn
’
t contain susceptible components to magnetic fields
Содержание MDCC-6430
Страница 1: ...ENABLING BRIGHT OUTCOMES MDCC 4330 MDCC 6430 User Guide Coronis Fusion 4MP 6MP...
Страница 5: ...5 K5902108 04 Coronis Fusion 4MP 6MP Welcome 1...
Страница 8: ...K5902108 04 Coronis Fusion 4MP 6MP 8 Welcome...
Страница 18: ...K5902108 04 Coronis Fusion 4MP 6MP 18 Display installation...
Страница 19: ...19 K5902108 04 Coronis Fusion 4MP 6MP Daily operation 3...
Страница 23: ...23 K5902108 04 Coronis Fusion 4MP 6MP Advanced operation 4...
Страница 36: ...K5902108 04 Coronis Fusion 4MP 6MP 36 Advanced operation...
Страница 37: ...37 K5902108 04 Coronis Fusion 4MP 6MP Cleaning your display 5...
Страница 39: ...39 K5902108 04 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Страница 42: ...K5902108 04 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Страница 43: ...43 K5902108 04 Coronis Fusion 4MP 6MP Important information 7...