Graphic board control panel
18
Disable RGB To Static Gray Color Translation
Select
Disable RGB To Static Gray Color Translation
if you wish to have
RGB values equally weighted, with 1/3 each. When not selected, then
the International Commission on Illumination (CIE – Commission
Internationale de I’Eclairage) model will be adapted, which weights
the colors as 59% Green, 30% Red, 11% Blue.
Monitor Configuration
The Monitor configuration page allows you to switch between the
following monitor configurations for the displays connected to a
BarcoMed controller:
• SingleView
A single large desktop spans across both displays.
• DualView
Both displays are enabled independent of each other.
Graphics Board
Update Firmware... Button:
Clicking this button will launch the
BarcoMed Hardware Configuration utility. This program allows the user
to flash update the firmware stored in the ROM of the currently
selected BarcoMed display controller. The BarcoMed Hardware
Configuration utility is implemented in a Wizard format, which guides
the user through the flash update procedure. The user will be
prompted to select a firmware update file to use for the update
process. This file will be provided by Barco Medical Imaging Systems if
and when a firmware update is required.
Device Details
From the
Device Details
tab you can access the Device Details page,
which provides detailed information about the display controller
connected to the active display. This information is useful in
debugging issues that may occur when using your MDMG-5121(s).