Display information
36
Reading room definition
This function is available only when the DICOM display function is
selected.
To modify a reading room definition:
1.
Select the DICOM display function. See
To select a display
function (DF):
, on page 33.
2.
In the ALC & DICOM Options menu, select
Reading Room Def
.
3.
Select the reading room definition you wish to modify. Thereto,
first select
Reading Room
. Select the appropriate reading room.
4.
To modify the
maximum ambient light
for this reading room,
select
Max. Ambient Light
and change the value.
You can change this value within certain limits determined by
the selected reading room.
5.
To modify the
Preset correction value
for this reading room,
select
Preset Corr. Value
and change the value.
You can change this value within certain limits determined by
the selected reading room.
Staff Office
Corresponds to light conditions in office
rooms.
Clinical Viewing
Corresponds to light conditions in diag-
nostic reading rooms for clinical view-
ing.
Emergency Room
Corresponds to light conditions in emer-
gency rooms.
Operating Room
Corresponds to light conditions in oper-
ating rooms. This setting has the high-
est maximum ambient light.
Name