Calibrating the Baron
After installation and before using the BARON, it has to be calibrated. This procedure has to be repeated every time the BARON is
disassembled and reassembled.
1 Press the ADJUST key to enter the adjustment mode, the adjustment menu will appear (menu 1).
2 Highlight INSTALLATION and press
ENTER
.
3 The installation menu is displayed on the screen.
4 Push the left and right cursor keys until the projection table is horizontal, check with water level (image 7.5).
5 Push the up and down keys until the horizontal tilted projection table reaches its highest position (image 7.5).
6 Press 2 times
ENTER
to store this reference position, the BARON is ready to be used.