Bante920 Benchtop pH/ORP Meter 

20 

 

pH Measurement 

1. 

Press MODE key until the meter shows

 

 

indicator. 

2. 

Rinse the pH electrode with distilled water to remove any impurities adhering to the probe body. 

3. 

Dip the electrode into the sample solution, stir the sensor gently.   

4. 

Wait for the reading to stabilize, record the measured value on the display. 

 

ORP Measurement 

Bante920 benchtop pH meter contains two millivolt measurement modes. 
 
ABSOLUTE MILLIVOLT MODE: 

 

Press MODE key until the display shows measurement unit “mV”, the meter is now enters absolute millivolt measurement mode. 

 

 

 
RELATIVE MILLIVOLT MODE: 

 

Press MODE key until the display shows

 

 

indicator, the meter enters relative millivolt measurement mode. 

 

 

 
Select one of the above modes. Dip the ORP electrode into sample solution, stir the sensor gently. Wait for the reading to stabilize, record the measured 
value on the display.   
 

Hold Function 

The meter contains two data hold modes. When the Auto-Hold function is enabled, the meter will automatically sense a stable endpoint reading and freeze it, 
“HOLD” indicator appears on the display. If the Auto-Hold function is disabled, press

 

 

key, the meter will immediately freeze currently displayed value. 

Press the key again to resume measuring. 

mV

Содержание Bante920

Страница 1: ...Bante920 Benchtop pH ORP Meter Instruction Manual BANTE INSTRUMENTS LIMITED...

Страница 2: ...conditions Relative humidity is less than 80 Ambient temperature is greater than 0o C and less than 60o C No potential electromagnetic interference The following list describes the standard accessori...

Страница 3: ...s meter is in the measurement mode Electrode slope indicator Indicates the average slope of the pH electrode Calibration mode indicator Indicates meter is in the calibration mode Stable indicator Indi...

Страница 4: ...and hold the key to enter temperature setting mode CAL Press the key to enter the calibration mode Press and hold the key to enter the setup menu MI Press the key to store current measured value Press...

Страница 5: ...necting the temperature probe 3 REF For connecting the reference electrode 4 USB For connecting the USB cable and DC5V power adapter Connecting the Electrode Holder The meter comes with an easy to use...

Страница 6: ...the screwdriver to tighten the screw moderately 3 Insert the plastic cover to previous position Installation is completed Connecting the Electrode 1 Take out the pH or ORP electrode from the packagin...

Страница 7: ...e cap from the bottom of the electrode FOR THE pH ELECTRODE If the glass sensitive membrane on the electrode dries out soak the sensor in 3M KCL solution or tap water for at least 15 minutes DO NOT us...

Страница 8: ...User Defined Buffer Calibration Points Select the number of calibration points you intend to calibrate the meter 1 point 2 points 3 points 4 points 5 points Resolution Sets the resolution for pH meas...

Страница 9: ...ithin a specified time period the meter will automatically show indicator 1 to 31 days Disable Date and Time Sets the date and time of the meter Clear stored data Clear all stored data Enable Disable...

Страница 10: ...ENTER key to confirm the display shows an option in the submenu 5 Press or key to select the desired option 6 Press ENTER key to confirm the meter returns to measurement mode Setting is completed EXI...

Страница 11: ...ollow the steps below to set the date and time during the first use 1 Press and hold the key for 3 seconds to enter the setup menu 2 Press or key until the display shows DATE option 3 Press ENTER key...

Страница 12: ...of temperature probe into corresponding connector socket Marked ATC The ATC indicator will show on the display the meter is now switched to automatic temperature compensation mode MANUAL TEMPERATURE...

Страница 13: ...ill not be accepted Make sure to calibrate the meter when attaching a new electrode or during first use DO NOT reuse calibration solution after calibration contaminants in solution will affect the cal...

Страница 14: ...ion 2 3 Rinse the pH electrode with distilled water Dip the electrode into the next standard buffer solution e g pH4 01 The meter automatically senses the current calibration solution and begins calib...

Страница 15: ...L2 The meter prompts you to continue with second point calibration 3 5 Rinse the pH electrode with distilled water Dip the electrode into the next custom buffer solution 3 6 Press or key to increase o...

Страница 16: ...trode for diagnostic purposes 1 Press MR key in the pH measurement mode the display shows LOC P 01 2 Press or key until the meter shows ELE P 02 Electrode Diagnosis 3 Press ENTER key to confirm the me...

Страница 17: ...in the meter shows the slope of each pH buffer group 6 After the browsing press MEAS key to exit the currect mode If you are not calibrate the meter or calibration is not successfully press MR key the...

Страница 18: ...DE key until the meter shows indicator 2 Rinse the ORP electrode with distilled water Dip the electrode into the calibration solution Stir the sensor gently 3 Press CAL key the meter shows the current...

Страница 19: ...key in the ORP measurement mode the display shows LOC P 01 2 Press or key until the meter shows ELE P 02 Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date For...

Страница 20: ...the temperature reading displayed differs from that of an accurate thermometer you need to calibrate the meter 1 Press and hold the o C key for 3 seconds to enter temperature calibration mode the disp...

Страница 21: ...e meter is now enters absolute millivolt measurement mode RELATIVE MILLIVOLT MODE Press MODE key until the display shows indicator the meter enters relative millivolt measurement mode Select one of th...

Страница 22: ...op pH meter allows up to 500 data sets to be stored and recalled MEMORY INPUT During the measurement process press MI key to input measured value into the memory Memory indicator appears on the displa...

Страница 23: ...nfirm the meter shows page number of the stored data 3 Press key the meter shows date and time of the stored data Format mm dd hh mm 4 Press key again the display shows the stored data 5 After the bro...

Страница 24: ...he stored data in meter will automatically send to computer INTERVAL RECORDING This function helps you to record the measuring value within the specify time period Click the Interval Recording button...

Страница 25: ...the electrode in a pH4 01 buffer solution for 10 to 30 minutes If this fails the electrode requires activation Soak the pH electrode in 0 1M HCl for 5 minutes Remove and rinse with deionized water th...

Страница 26: ...ent Cosmetics Dairy Products Education Fats Cream Field Use Fish Products Lab Flasks Low Ionic Meat Cheese Micro Samples Paint Photographic Soil Surface Test Tubes Tris Buffer Viscose Samples Troubles...

Страница 27: ...ts 1 point Only for Relative mV mode Calibration Range 200mV Temperature Range 0 105 C 32 221 F Accuracy 0 5 C 0 9 F Resolution 0 1 C Calibration Points 1 point Calibration Range Measured value 10 C O...

Страница 28: ...w to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover sensor and calibration solutions Out of warranty products will be...

Отзывы: