background image

   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Bante321 Portable Ion Meter 

Instruction Manual 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
BANTE INSTRUMENTS CO., LTD 

 

Содержание Bante321

Страница 1: ...Bante321 Portable Ion Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Страница 2: ...urrent work environment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following...

Страница 3: ...con Indicates the measuring value has stabilized Calibration mode icon Indicates the meter is in the calibration mode Hold icon Indicates the measuring value has been locked Setup mode icon Indicates...

Страница 4: ...rs the setup menu Press and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report or data logs Decrease value or sc...

Страница 5: ...the DC5V power adapter with USB cable for power supply NOTE take out the batteries Connecting the Sensors Take out the ion selective electrode from the packaging Insert the connector into the BNC conn...

Страница 6: ...he displayed option to meet measurement requirements The following table describes the functions of the menu items INDEX MENU DESCRIPTION OPTIONS DESCRIPTION DEFAULT Measurement Unit Set the ion conce...

Страница 7: ...nd Time Set the current date and time Year month day hour minutes Clear Stored Data Delete all stored readings in the memory Enable Disable Disable Factory Reset If enabled all of the calibration data...

Страница 8: ...TE P 08 2 2 Press the Enter key the meter shows current year e g 2018 2 3 Press the or key to set year and the Enter key to confirm the display shows current date and time Format month day hour minute...

Страница 9: ...w switched to the automatic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the C icon will show on the display indicating that the meter...

Страница 10: ...d or calibration is not successfully the display will always show 0 000 During the calibration ensure that the selected calibration points cover the anticipated range of the samples 1 1 Press the Mode...

Страница 11: ...electrode with distilled water then rinse with a small amount of standard solution Place the electrode into the next standard solution Stir the electrode gently Press the Enter key the Calibration ic...

Страница 12: ...key until the meter shows ELE P 02 2 3 Press the Enter key the meter shows the last calibration date Format month day 2 4 Press the key the meter shows the first calibartion point and mV value e g 100...

Страница 13: ...s contain the interference ions adding the ionic strength adjuster to sample and using the plastic beaker are necessary 1 Press the Mode key until the display shows icon 2 Rinse the ion selective elec...

Страница 14: ...stored readings 1 Press the MR key in the measurement mode the meter shows LOC P 01 Data Log 2 Press the Enter key the meter shows the serial number of the stored data 3 Press the key the meter shows...

Страница 15: ...lick the OK button to confirm Click the Receive button the stored data automatically transfer to computer Interval recording This function is used for recording the measuring value within the specify...

Страница 16: ...ation Electrode has expired Replace the electrode Keypad is not working properly Replace the batteries Specifications Ion Model Bante321 Range 0 001 19999ppm mg L mol L Depending on the range of ISE A...

Страница 17: ...ppm Ag US Silver Ag 0 01 107900ppm S US Sulphide S2 0 003 32100ppm NH3 US Ammonia NH3 0 02 17000ppm Addendum 2 Preparation of Ion Standard Solutions 1000ppm To prepare these solutions half fill a 1 li...

Страница 18: ...ride 1 91g Silver Ag Silver Nitrate 1 57g Sulphide S2 Sodium Sulfide 7 49g Ammonia NH3 Ammonium Chloride 3 82g Swirl the flask gently to dissolve the reagent and fill to the mark with distilled water...

Страница 19: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Отзывы: