Bante321 Portable Ion Meter
11
Viewing the ion calibration report
2.1
Press the
MR
key in the ion measurement mode, the meter shows LOC/P-01.
2.2
Press the
or
key until the meter shows ELE/P-02.
2.3
Press the
Enter
key, the meter shows the last calibration date (Format: month-day).
2.4
Press the
key, the meter shows the first calibartion point and mV value (e.g., 100ppm,
59mV).
2.5
Press the
key, the meter shows the second calibartion point and mV value (e.g., 1000ppm,
118mV).
2.6
To exit the calibration report, press the
Meas
key.