Bante321 Portable Ion Meter
10
1.5
Rinse
the
ion
selective electrode with
distilled
water, then
rinse
with
a
small
amount of
standard solution. Place the electrode into corresponding standard solution.
Stir the electrode
gently to create a homogeneous solution. Press the
Enter
key, the Calibration icon begins
flashing.
1.6
Wait for the reading to stabilize, the display shows 1000ppm/CAL2. The meter prompts you to
continue with second point calibration.
1.7
Rinse the ion selective electrode with distilled water, then rinse with a small amount of
standard solution. Place the electrode into the next standard solution. Stir the electrode
gently. Press the
Enter
key, the Calibration icon begins flashing.
1.8
Wait for the reading to stabilize, the meter automatically show END and return to the
measurement mode. Calibration is completed.
•
If you have selected the multi-point calibration in the setup menu, the display will show CAL3.
The meter prompts you to continue with third point calibration. Repeat the step 1.7 above
until the display shows END. The meter will automatically return to the measurement mode.
•
If you want to exit the calibration, press the
Meas
key.