ECscan10 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Electrode   
Maintenance and Replacement 

  Rinse the electrode thoroughly with distilled water after use. 

  Do not touch the platinum black coating on the sensor surface and 

  always keep it clean. 

  If there is a build-up of solids inside the sensor, remove carefully, 

  then recalibrate the tester. 

  If you do not use the tester for long periods, store the electrode 

  with tap water. 
 

Replacing the Electrode 

If the tester fails to calibrate or gives fluctuating readings, you should 
consider replacing the electrode. 
1.  Twist  the  electrode collar counter  clockwise, pull the electrode 
  away from the tester. 

 

 

2.  Align the slot on the new electrode, gently push the electrode into 
  the tester. 

 

 

3.  Twist the electrode collar clockwise until tight. 

 

 

 

 

 

 

Appendix 

 

Preparation of Conductivity Standard Solutions 

Place the analytical grade potassium chloride (KCl) in a beaker and 
dry in an oven for about 3 hours at 105° C (221° F), then cool to room   
temperature. Add the reagent to a 1 liter volumetric flask according 
to the instructions in table below. Fill the distilled water to the mark, 
mix the solution until the reagent is completely dissolved. 

 
Conductivity Standard 

Reagent 

Weight 

146.5 µ S/cm 

KCl 

74.4 mg 

1413 µ S/cm 

KCl 

745.5 mg 

12.88 mS/cm 

KCl 

7.45 g 

 

 

Optional Accessories 

 
Order Code 

Description 

E-ECscan-C1-100K 

Platinum conductivity cell, K=1 

ECCS-146 

Standard solution 146.5

 

μ

S/cm, 480

 

ml 

ECCS-1413 

Standard solution 1413

 

μ

S/cm, 480

 

ml 

ECCS-1288 

Standard solution 12.88

 

mS/cm, 480

 

ml 

 

Tester Specifications 

 
Model 

ECscan10 

Conductivity 

 

Range 

ECscan10

 

L:  1.0 to 199.9

 

μ

S/cm 

ECscan10

 

M:  10 to 1999

 

μ

S/cm 

ECscan10

 

H:  0.1 to 19.99

 

mS/cm 

Resolution 

0.1, 1 

Accuracy 

± 1% F.S 

Calibration Point 

1 point 

Calibration Solution 

ECscan10

 

L:  146.5

 

μ

S/cm 

ECscan10

 

M:  1413

 

μ

S/cm 

ECscan10

 

H:  12.88

 

mS/cm 

Temperature Compensation  0 to 50° C, automatic 
Temperature 

 

Range 

0 to 50° C 

Resolution 

1° C 

Accuracy 

± 1° C 

Other Specifications 

 

Cell Constant 

K=1 

Operating Temperature 

0 to 50° C (32 to 122° F) 

Storage Temperature 

0 to 60° C (32 to 140° F) 

Relative Humidity 

< 80% (non-condensing) 

IP Rating 

IP54 

Display 

LCD, 21 ×  21 mm (0.82 ×  0.82 in.) 

Power Requirements 

3 ×  1.5V LR44 micro alkaline batteries 

Auto-Off 

8 minutes after last key pressed 

Dimensions 

185 (L) ×  40 (Dia.) mm (7.28 ×  1.57 in.) 

Weight 

100

 

g (3.5 oz.) 

 