Overview 

Thank you for selecting the ECscan series pocket conductivity tester. 
This user

 

manual provides a step-by-step guide to help you operate

.

 

the tester, please carefully read the following instructions before use. 
 

Installing the Batteries 

1.  Remove the battery compartment lid. 

 

 

2.  Insert three LR44 alkaline batteries into the battery compartment, 
  note polarity. 

 

 

3.  Replace the battery compartment lid to its original position and
  turn clockwise until tight. 

 

 

Keypad 

 
Key     

Function 

 

  Switch the tester on or off 

  Lock or unlock measurement 

  Confirm or exit the calibration mode 

  Return to the conductivity measurement mode 

 

  Press and hold the key to start the calibration 

  Increase value during the setting 

 

  Show the temperature readings 

  Decrease value during the setting 

 
 

 

 

Prior to Use 

Remove the protective cap and translucent cover from the bottom of 
the tester. If the platinum sensor has dried out, soak the electrode for 
about 10 minutes in tap water. 

 

 

 

 

Switching the Tester On and Off 

  Press the 

Meas

 key to switch on the tester. 

  Press and hold the 

Meas

 key to switch off the tester. 

 

 

 

 

If you do not press any key within 8 minutes, the tester will switch off 
automatically to conserve energy. 
 

 

 

Conductivity Calibration 

The ECscan10 conductivity tester allows 1 point calibration. For better 
accuracy, we recommend to calibrate the tester regularly. The following 
table shows acceptable standard solutions for each tester. 
 
Model 

Standard Solution Range   

Default 

ECscan10

 

100 to 180 µ S/cm 

146.5 µ S/cm 

ECscan10

 

1000 to 1800 µ S/cm 

1413 µ S/cm 

ECscan10

 

10 to 18 mS/cm 

12.88 mS/cm 

 
Make sure that using a fresh standard solution during the calibration. 
DO NOT reuse the standard solution after calibration, contaminants 
in solution will affect the calibration and eventually the accuracy of 
the measurement. 

 

1.  Rinse the electrode with distilled water and place into the standard 
  solution, stir tester gently to remove air bubbles trapped in the slot 
  of the sensor. 

Tap water 