ECscan10
I
2
2. Press and hold the
Cal
key to enter the calibration mode, the
display shows ---.
3. Press the
/
key to set the calibration value, make sure that
the displayed value matches the calibration standard.
4. Press the
Meas
key, the calibration value will automatically flash
three times.
5. Press the
Meas
key again, the tester returns to the measurement
mode, the CAL icon disappears from the display. Calibration is
completed.
•
During the calibration, if the display shows
indicating that
the measured conductivity value deviates from the theoretical
value of the standard solution. The calibration will not be accepted.
Please replace the fresh standard solution and calibrate the tester
again.
•
During the setting, press and hold the
/
key will make the
value change faster.
•
If you want to exit the calibration, DO NOT press the
Meas
key in
the step 4. Press and hold the
/
key until the display shows
---, press the
Meas
key to exit.
Measurement
Conductivity Measurement
Rinse the electrode with distilled water, place the electrode into the
sample solution and stir gently, make sure that no air bubbles on the
sensor surface. Wait for the measurement to stabilize and record the
reading.
•
During the measurement, DO NOT completely immerse the tester
in water.
•
Press the key, the tester will lock the measurement, the HOLD
icon appears on the display. Press the key again to resume
measurement.
•
If the display shows indicating the measurement exceeds the
range, remove the tester from the sample immediately.
Temperature Measurement
•
Press the
o
C
key, the tester shows the temperature readings.
•
Press the
Meas
key to return to the conductivity measurement.
Platinum sensor