DOscan10   

I

    4 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

2 Point Calibration 

Ensure that you have selected 2 points calibration in the setup menu. 
2.1  Repeat steps 1.1 and 1.2 above. When the first calibration point

.

 

 

is completed, the display will show 0%/

 

CAL2, the tester prompts

.

 

 

you to continue with second point calibration. 

 

 

 
2.2  Place the dissolved oxygen electrode into the zero oxygen solution

.

 

 

for about 10 minutes, press the 

Enter

 key to begin the calibration. 

 

When the reading has stabilized, the tester will show

 

.   

 

Calibration is completed. 

 

 

 

 

 

  Performing a percentage saturation calibration will calibrate the 

 

corresponding mg/L or ppm concentration value simultaneously. 

  To exit the calibration without saving calibrated values, press the 

 

Meas

 key. 

 

 

 

Temperature Calibration 

1.  Press and hold the 

Cal

 key to enter the setup menu. 

 

 

 
2.  Press the 

Enter

 key until the display shows

 

/

 

 

or

 

/

 

 

 

 

 
3.  Press the 

Cal

 key, the tester enters the temperature calibration 

 

mode. 

 

 

 
4.  Place the electrode into a solution with a known accurate   
 

temperature and wait for the measurement to stabilize. 

 

 

 

 
5.  Press the 

Cal

 key to modify the temperature value. 

 

 

 

 
6.  Press the 

Enter

 key to save and press the 

Meas

 key to return 

 

to the measurement mode. 

 

 

 

 

 

 

Measurement 

The DOscan10 tester can be used to measure the water, wastewater, 
brine and other liquids. If your sample is seawater or water containing 
large amounts of salt, make sure to set the salinity coefficient before 
measurement. Some gas and steam such as chloride, sulfur dioxide, 
sulfureted hydrogen and carbon dioxide can permeate the membrane 
via diffusion. Their existence will influence the measurements. If the 
sample contains solvent, grease, sulfide and alga, the membrane will 
be damaged or eroded. 
 

Содержание DOscan10

Страница 1: ...DOscan10 Pocket Dissolved Oxygen Tester USER MANUAL Bante Instruments Inc ...

Страница 2: ...the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use Filling the Electrolyte Solution 1 1 Take out the dissolved oxygen electrode from the carrying ...

Страница 3: ...rement endpoint Enable Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Disable default Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode ...

Страница 4: ...d the single point calibration in the setup menu we recommend that you perform a 100 saturation calibration in the air saturated water If the 2 points calibration is selected the zero oxygen solution needs to be used During the calibration and measurement the temperature sensor on electrode must be immersed in the sample solution completely the solution should keep 0 3 m s of minimum flow rate to ...

Страница 5: ... hold the Cal key to enter the setup menu 2 Press the Enter key until the display shows or 3 Press the Cal key the tester enters the temperature calibration mode 4 Place the electrode into a solution with a known accurate temperature and wait for the measurement to stabilize 5 Press the Cal key to modify the temperature value 6 Press the Enter key to save and press the Meas key to return to the me...

Страница 6: ... cap is contaminated or the electrolyte solution is depleted Screen shows Electrode is broken Replace the dissolved oxygen electrode 5 I DOscan10 1 Set the barometric pressure and salinity coefficient in the setup menu 2 Rinse the electrode with distilled water 3 Place the electrode into the sample solution and stir gently Wait for the measurement to stabilize and record the reading During the mea...

Страница 7: ...ster is one year from the date of shipment Above warranty does not cover the electrode and electrolyte solution Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mo...

Отзывы: