background image

3   

I   

DOscan10 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Table 1. Altitude and Barometric Pressure 

 
Altitude (m) 

 

 

kPa 

mmHg 

 

 

101.3 

760 

100 

 

 

100.1 

750 

200 

 

 

98.8 

741 

300 

 

 

97.6 

732 

400 

 

 

96.4 

723 

500 

 

 

95.2 

714 

600 

 

 

94.0 

705 

700 

 

 

92.8 

696 

800 

 

 

91.7 

688 

900 

 

 

90.5 

679 

1000 

 

 

89.4 

671 

1100 

 

 

88.3 

662 

1200 

 

 

87.2 

654 

1300 

 

 

86.1 

646 

1400 

 

 

85.0 

638 

1500 

 

 

84.0 

630 

1600 

 

 

82.9 

622 

1700 

 

 

81.9 

614 

1800 

 

 

80.9 

607 

1900 

 

 

79.9 

599 

2000 

 

 

78.9 

592 

2100 

 

 

77.9 

584 

2200 

 

 

76.9 

577 

2300 

 

 

76.0 

570 

2400 

 

 

75.0 

563 

2500 

 

 

74.1 

556 

2600 

 

 

73.2 

549 

2700 

 

 

72.3 

542 

2800 

 

 

71.4 

536 

2900 

 

 

70.5 

529 

3000 

 

 

69.6 

522 

3100 

 

 

68.7 

515 

3200 

 

 

67.9 

509 

3300 

 

 

67.0 

502 

3400 

 

 

66.2 

496 

3500 

 

 

65.4 

490 

3600 

 

 

64.6 

484 

 

 

 

Dissolved Oxygen Calibration 

The DOscan10 tester allows 1 or 2 points dissolved oxygen calibration. 
If you have selected the single point calibration in the setup menu, 
we recommend that you perform a 100% saturation calibration in the

.

 

air-saturated water. If the 2 points calibration is selected, the zero

.

 

oxygen solution needs to be used. 
 

 

During the calibration and measurement, the temperature sensor  on 
electrode must be immersed in the sample solution completely, the 
solution should keep 0.3 m/s of minimum flow rate to avoid oxygen 
starvation at the membrane. 
 

 

 

 

Single Point Calibration - 100% Saturation 

Ensure that you have selected 1 point calibration in the setup menu. 
1.1  Press the 

Cal

 key, the display shows 100%/

 

CAL1. 

 

 

 
1.2  Hold the dissolved oxygen electrode in the air at 100% relative 
 

humidity or place the electrode into the air-saturated water for 

 

about 10 minutes. Press the 

Enter

 key to begin the calibration. 

 

 

 

 
1.3  When the reading has stabilized, the tester will show

 

 

and 

 

return to the measurement mode. 

 

 

Temperature sensor

 

Содержание DOscan10

Страница 1: ...DOscan10 Pocket Dissolved Oxygen Tester USER MANUAL Bante Instruments Inc ...

Страница 2: ...the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use Filling the Electrolyte Solution 1 1 Take out the dissolved oxygen electrode from the carrying ...

Страница 3: ...rement endpoint Enable Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Disable default Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode ...

Страница 4: ...d the single point calibration in the setup menu we recommend that you perform a 100 saturation calibration in the air saturated water If the 2 points calibration is selected the zero oxygen solution needs to be used During the calibration and measurement the temperature sensor on electrode must be immersed in the sample solution completely the solution should keep 0 3 m s of minimum flow rate to ...

Страница 5: ... hold the Cal key to enter the setup menu 2 Press the Enter key until the display shows or 3 Press the Cal key the tester enters the temperature calibration mode 4 Place the electrode into a solution with a known accurate temperature and wait for the measurement to stabilize 5 Press the Cal key to modify the temperature value 6 Press the Enter key to save and press the Meas key to return to the me...

Страница 6: ... cap is contaminated or the electrolyte solution is depleted Screen shows Electrode is broken Replace the dissolved oxygen electrode 5 I DOscan10 1 Set the barometric pressure and salinity coefficient in the setup menu 2 Rinse the electrode with distilled water 3 Place the electrode into the sample solution and stir gently Wait for the measurement to stabilize and record the reading During the mea...

Страница 7: ...ster is one year from the date of shipment Above warranty does not cover the electrode and electrolyte solution Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mo...

Отзывы: