A180/181
I
38
If you need to set the Total Volume and Sample Volume, press the
/
key until the option shows "ON". Press the
Enter
key, the setting
value will show on the screen.
2.6 Press the
/
key to set the parameter, press the
Enter
key
to confirm. When the setting is completed, the meter will show
calculated BOD result.
2.7 Press the
ESC
key, the meter will take a new measurement.
If you want to exit the BOD measurement, press the
Mode
key to
convert the measurement mode.
OUR/SOUR Measurement
The A180 meter contains an application for the calculation of Oxygen
Uptake Rate (OUR) and Specific Oxygen Uptake Rate (SOUR). Before
the measurement, make sure that the OUR or SOUR parameters have
been set to desired values.
OUR/SOUR Parameters
Sample ID
Set the sample ID to associate reading with the data log.
0000 to 9999
Default 0000
Sample Volume
Set the sample volume.
Default 100 ml
Total Volume
Set the total volume of sample.
Default 100 ml
Minimum Testing Time
Set the minimum time of measurement. When the minimum time is
reached, the measurement will start.
Default 1 minute
Maximum Testing Time
Set the maximum time of measurement. When the maximum time is
reached, the measurement will end.
Default 5 minutes
Minimum Beginning DO
Set the minimum dissolved oxygen value allowed at the start of the
measurement.
Default 0.00 mg/L
Minimum Ending DO
Set the minimum dissolved oxygen value allowed during the
measurement. If the measured value falls below this value, the
measurement will end.
Default 0.00 mg/L
Solids Weight
Set the total solids or volatile suspended solids concentration of the
sample.
Default 1.00 g/L
SOUR@20° C
This option is applied to the SOUR measurement only. If enabled, the
readings will automatically reference to 20° C.
Enable
Disable
Default