A130/131
I
20
2.2 Press the
/
key to select an electrode type for storing and
recalling the electrode slope, press
Enter
key to confirm.
The A series meter is able to store up to 3 electrode slopes in
memory. For example, you select the
Ion 1
and use the fluoride
electrode to calibrate the meter, select the
Ion 2
and use the
chloride electrode to calibrate the meter. The electrode slopes
will be saved in the selected electrode type separately after the
calibration.
Ion Settings
The A series meter contains 7 ion settings and 10 general settings in
the setup menu.
Menu Items and Options
Sample ID
Set the sample ID to associate reading with the data log.
0000 to 9999
Default 0000
Concentration Unit
Set the default ion concentration unit.
ppm
Default
mg/L
mol/L
mmol/L
Water Hardness Unit
Set the default water hardness unit.
° dH (German degree)
Default
° e (English degree)
° f (French degree)
mg/L (CaCO
3
), mg/L (CaO), mg/L (Ca
2+
)
mmol/L
Calibration Points
Set the number of calibration points.
2 to 5 points
Default 2 points
Ionic Valency
Set the valence of the ion selective electrode.
Monovalent
Default
Divalent
Alarm Limits
Set the high and low limit values to activate alarm.
Enable (setting range: 0 to 30000)
Disable
Default
Calibration Due
Set the calibration interval to activate alarm.
Enable (setting range: 1 to 31 days)
Disable
Default
If you want to change the current settings, press and hold the
key
to enter the setup menu, press the
/
key to select an option or
set a value, press the
Enter
key to confirm.
If the ion concentration unit has converted, the meter will show "Can
not find the electrode slope" always and wait for calibration. Press
the
Cal
key and refer to the
Ion Calibration
section to perform the
calibration, the meter will switch to selected unit when calibration is
completed.
•
During the setting process, press and hold the
/
key will
make the value change faster.
•
To exit the setting without saving changes, press the
ESC
key.