background image

P/N 64118 rev. B

Control Module Configuration

22

Banner Engineering Corp. 

• 

Minneapolis, U.S.A.

www.bannerengineering.com  •  Tel: 763.544.3164

Figure 5-9.   Use the PSF Configuration screen to program each control module individually

Serial Transmission

Specifies the type of data transmitted 

from the control module to its host  

after each scan.

Measurement Mode Result:

 Data 

transmitted will reflect the Analysis Mode 

selections.

All Mode:

 Transmits all data.

Max. Meas. Mode: 

Sends only the largest 

measurement in each measuring event, to 

decrease transmission size and speed 

response. Choose to send when the array 

is clear or send at the host’s request.

Transmission Type: 

ASCII or Binary, 

defines the format in which the data 

will be sent.

Serial Options:

 Suppress Clear Data 

or Suppress Header to decrease 

transmission size and speed response.

Analysis (Measurement)

Mode Selection

Choose the measurement  

option that best tells you  

the size and/or position 

of objects as they

relate to the array.

Control Mode Selection

Continuous Mode:

 The control 

module constantly polls  

the array for status.

Host Mode:

 The control module  

polls the array for status when  

prompted by a host controller.

Gate Mode: 

The control module 

polls the array for status when 

prompted by an input from a  

Gate sensor.

Scanning Method

Straight scan

 polls each beam 

sequentially to determine the target 

object’s overall size. This is the most 

accurate and precise measurement, 

but also the most time-consuming.

Single Edge scan

 requires the target 

object to block beam 1 (closest to 

the sensors’ cabled ends), then 

conducts a time-saving binary 

search to “hunt” for the target’s 

overall height (one variable edge). 

Double Edge scan

 conducts a 

time-saving search of the entire 

array to “hunt” for the target’s 

overall width (two variable edges).

Set Point and 

Hysteresis Selection

Assigns the set point to 

determine where within the 

array the output(s) will 

respond and hysteresis 

values to smooth output 

response.

Trigger/Trigger Channel Number

May be used to trigger (or gate) the scan sequence of another 

A-GAGE High-Resolution MINI-ARRAY

 controller; in straight scanning 

mode, it defines when during each scan discrete Output #2 will change 

state.

Scan #: 

(1-9) Analog outputs are 

updated with an average value of the 

data received during the selected 

number of scans; discrete outputs 

respond only if the received data is 

identical for all of the selected number of 

consecutive scans.

Analog and Discrete

Output Assignment

Assigns an analysis 

(measurement) mode 

to each output.

Alarm:

 Causes the control module to 

turn on discrete Output #2 whenever 

the System detects a sensing error or 

if the optical signal becomes marginal.

Selected Controller 

Identifies the specific 

control module being 

configured.

Serial Communication

Changes the identification 

and baud rate of the

controller being

configured.

Zero Value 

Zero Value is used to specify the 

analog output when the 

measurement mode value is zero. 

The user can specify a value 

of LAST, NULL, or SPAN.

 

LAST:

 

 Output holds the last non-zero value before the 

light screen became clear.

 

NULL:

  Provides the minimum value.

 

SPAN:

 

Provides the maximum value.

Содержание A-GAGE High-Resolution MINI-ARRAY MAHCIN-1

Страница 1: ...Alignment routine equalizes gain of each beam for reliable 2 5 mm 0 10 object detection throughout the array Host computer or PLC may be used to initiate scans and or process scan data Unique addresse...

Страница 2: ...P N 64118 rev B 2 Banner Engineering Corp Minneapolis U S A www bannerengineering com Tel 763 544 3164...

Страница 3: ...8 5 2 2 Software Alignment Routine 19 5 2 3 Blanking 20 5 3 Programming Control Module Response 21 5 3 1 Selected Controller and Serial Connection 23 5 3 2 Control Mode Selection 23 5 3 3 Scanning Met...

Страница 4: ...System configuration A host computer or PLC may be used to control and or receive input from the System Sensors are available in 12 array lengths from 163 mm to 1951 mm 6 4 to 76 8 in 163 mm 6 4 incre...

Страница 5: ...2 256 320 384 448 512 576 704 Emitter and Receiver Models Cables Configure and monitor the System with the supplied software and any PC compatible computer running Windows XP Vista or 7 via an RS 232...

Страница 6: ...ion 5 3 5 for more information on outputs Alarm output discrete output 2 conducting and gate signal received a green LED signals that the sensors are properly aligned A segmented LED Diagnostics Indic...

Страница 7: ...patible computer running Windows The menu driven program walks the user through the many scanning and output options After the desired options are selected the user can save the combination of selecti...

Страница 8: ...indicates sensors misaligned or one or more beam s blocked Connections Sensors connect to controller using two 5 conductor quick disconnect cables one each for emitter and receiver ordered separately...

Страница 9: ...MAHR26A Receiver 724 mm 28 5 756 mm 29 7 MAHE32A Emitter MAHR32A Receiver 887 mm 34 9 918 mm 36 2 MAHE38A Emitter MAHR38A Receiver 1049 mm 41 3 MAHE45A Emitter MAHR45A Receiver 1215 mm 47 8 1246 mm 49...

Страница 10: ...1V 10 mA and 1 5V 150 mA Serial Data Outputs RS 232 or RS 485 interface Up to 15 control modules may be given unique addresses on one RS 485 party line ASCII or binary data format 9600 19 2K or 38 4K...

Страница 11: ...4118 rev B 81 0 mm 3 19 115 0 mm 4 53 106 0 mm 4 17 35 0 mm 1 38 DIN mounting slot Slot for M3 5 screws 2 5 5 mm 0 22 69 0 mm 2 72 5 0 mm 0 20 96 0 mm 3 78 Figure 2 2 Control module dimensions Specifi...

Страница 12: ...vibration isolators and M4 Keps nuts all supplied See Figure 3 1 Standard M4 or 8 32 bolts may be substituted and the vibration isolators eliminated in situations where the emitter and receiver are no...

Страница 13: ...st point in the same direction Measure from one or more reference planes e g the building floor to the same point s on the emitter and receiver to verify their mechanical alignment If the sensors are...

Страница 14: ...x 0 10V ANALOG OUTPUT 2 25 mA max 0 10V ANALOG OUTPUT 1 DISCRETE OUTPUT 2 ALARM 150 mA max Figure 3 4 MAHCVP 1 hookup 30V dc max POWER 1 16 30V dc 1 A max 10 30V dc GATE SIGNAL 30V dc 2 3 4 5 6 7 8 9...

Страница 15: ...e open collector output transistor of a dc sensing device The gate signal controls scanning when one of four Gate options is selected in the Control Mode Selection menu of the PSF configuration routin...

Страница 16: ...engineering com as required 2 Insert the Software CD into the CD drive If the program does not auto start browse to your CD drive click Setup exe then select START then select RUN 2 The Welcome dialog...

Страница 17: ...m supports serial communication via the COM1 COM20 port of the computer Highlight the desired serial port to select it a bullet will appear next to the selected option then select Save Settings on Exi...

Страница 18: ...ervice the Diagnostics Indicator will show or see Figure 5 4 5 2 1 Push button Alignment Routine At installation or whenever the emitter and or receiver is moved the System should be re aligned Press...

Страница 19: ...ore Control module Settings Auto Blanking Abort Auto Blanking Save to File Read From File Cancel OK and Edit To access any of these sub routines first press Stop then the selected option Start command...

Страница 20: ...n to Run mode again press the Alignment switch for 3 seconds When the System is ready for operation and configured for beams to be blanked the period on the Diagnostics Indicator will remain lit showi...

Страница 21: ...ed for beams to be blanked the period on the Diagnostics Indicator will remain lit showing see Figure 5 4 Figure 5 8 Edit channel blank state screen showing beam 22 and beams 35 42 blanked beams 65 80...

Страница 22: ...ngle Edge scan requires the target object to block beam 1 closest to the sensors cabled ends then conducts a time saving binary search to hunt for the target s overall height one variable edge Double...

Страница 23: ...an control method it is used in most analog output applications and whenever continuous updating of the outputs is acceptable Host Mode allows the control module to communicate with a host computer or...

Страница 24: ...e Figure 5 13 4 This routine continues to narrow the field until the edge is found Note that the receiver always checks the bottom beam first and only if that beam is blocked does the binary search co...

Страница 25: ...provide a solid obstruction the size of the object will determine the maximum step size Figure 5 14 Sensor response time is a function of sensor length and scanning method Typical scan times are show...

Страница 26: ...ssigned to any one of the available outputs see Figure 5 18 Each output can be set for MEAS1 MEAS2 MEAS1 Inverted or MEAS2 Inverted 5 3 5 Analog Output Configuration Analysis Mode Assignment Analog ou...

Страница 27: ...values will be at mid range 5 3 6 Zero Value Zero value is used to specify the analog output when the measurement mode value is zero The user can select an analog output of LAST NULL or SPAN LAST Outp...

Страница 28: ...at each set point High and Low to cause the associated output to change state Hysteresis prevents unstable output conditions when the scan analysis value exactly matches one of the set points The defa...

Страница 29: ...two Scan Analysis Mode selections can be active ALL Mode the status of each beam in the light screen array will be transmitted for each scan In ALL mode blanked channels are transmitted as unblocked...

Страница 30: ...t transmission of the three byte identification string header string associated with either ASCII or Binary data formats The header string consists of two bytes at the start of the data and a terminat...

Страница 31: ...ponse to the currently loaded PSF select Execute Figure 5 26 The Measurement Output screen will appear Figure 5 29 select Run The table will display the status for the selected measurement mode s incl...

Страница 32: ...n Receiver DIN Rail Mountable Control Module HIGH RESOLUTION MINI ARRAY CONTROLLER ALIGNMENT SWITCH POWER POWER DIAGNOSTICS INDICATOR 2 TX 3 RX 5 COM RS 232 MAHCVP 1 OUTPUT ALARM GATE ALIGN 1 16 30V d...

Страница 33: ...t array alignment is satisfactory To improve alignment adjust the position of the emitter and receiver until the green LED lights making sure that no unblanked beams are interrupted Then use one of th...

Страница 34: ...ter A msg 2 83 assume initiation byte which is the ASCII letter S The host transmits this three byte message at the defined baud rate The format is one start bit one stop bit even parity and eight dat...

Страница 35: ...the control module sends the status of a clear condition only on the first clear scan After that the control module will continue to scan but will not transmit data until the sensor is again blocked...

Страница 36: ...Meas Mode causes the control module to retain the maximum measurement values for a given object These values are transmitted either immediately after the control module is cleared or when prompted by...

Страница 37: ...37 Banner Engineering Corp Minneapolis U S A www bannerengineering com Tel 763 544 3164 P N 64118 rev B Banner Engineering Corp Minneapolis U S A www bannerengineering com Tel 763 544 3164...

Страница 38: ...unobstructed from the emitter to the receiver same as a made or unblocked beam Excess Gain A measurement of the amount of light falling on the receiver from the emitter over and above the minimum amo...

Страница 39: ...39 Banner Engineering Corp Minneapolis U S A www bannerengineering com Tel 763 544 3164 P N 64118 rev B Banner Engineering Corp Minneapolis U S A www bannerengineering com Tel 763 544 3164...

Страница 40: ...ANTIES WHETHER EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE AND WHETHER ARISING UNDER COURSE OF PERFORMANCE COURSE OF DEALING OR...

Отзывы: