5/
Operation
20/38
1786-001 en/2022-08
Fig. 5
Fill oscillating tank
1 Turning knob to adjust the heating temperature
2 Filling level mark
3 Drain with ball valve (depending on model)
4 Turning knob to set the ultrasound duration
Requirements
• For ultrasound baths with an outlet, the ball valve must be
closed.
• The ultrasound and heating must be turned off.
Procedure
1. Fill the ultrasonic oscillating tank with water up to
1
⁄
3
.
2. Pour the correct quantity of agent into the ultrasonic oscillating
tank.
3. Fill the tank with water up to the filling level mark while avoiding
foam formation. When using indirect sonication, take the dis-
placement caused by inset vessels into consideration.
Result
» The device is ready to switch on.
15
13
10
9
7
5
0
3
1 min
80
70
60
50
40
30
°C
2
4
3
1