Safety
/2
2.10 Preventing damage to the ultrasound bath
• Only use aggressive preparations in inset beakers or insert tubs.
When working with aggressive preparations, avoid splashing in
the contact liquid or on the stainless steel surface. Immediately
replace contaminated sonication fluid. Clean surfaces and wipe
them dry.
• If strongly acidic preparations are used, the ball of the ball valve
can become corroded. The ball valve will leak. If the use of a
strongly acidic detergent cannot be avoided, use a stainless
steel ball valve.
• Do not operate the ultrasound bath without sonication fluid in
the oscillating tank. Make especially sure that the heating is
switched off when the oscillating tank is empty. The fill level
must always be at or slightly above the filling level mark.
2.11 Disturbance of wireless communication
The ultrasound bath may interfere with other wireless communica-
tion ultrasound baths in the immediate vicinity, such as:
• mobile phones,
• WLAN devices,
• Bluetooth devices.
If interference occurs with the operation of a wireless device, move
it further away from the ultrasound bath.
The ultrasound bath meets the requirements for class B devices
according to EN 55011.
2.12 Safety stickers on the ultrasound bath
• Observe all safety labels on the ultrasound bath.
• Keep the safety stickers in legible condition. Do not remove
them. Replace them if they are no longer legible. To do so,
please contact our Customer Service. See chapter
.