Ultrasonic converter
Step/booster horn
Vigorously turn sickle spanner
HS 40/42 long, in the opposite
direction
30 / 42
51376-012 en/2022-09
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.4
Dismounting and mounting the standard/booster horns
Standard/booster horns
• are screwed onto the respective ultrasonic converter,
• are made of high-strength titanium alloy,
• transmit the oscillations to the probe,
• enhance the amplitude thanks to their geometry.
At the time of delivery, standard horn SH 100 G is
fi
rmly mounted onto the UW 100, and booster
horn SH 200 G is
fi
rmly mounted onto the UW 200.
Before mounting the standard/booster horns, the ultrasound generator must be turned o
ff
and the ultrasonic converter must be disconnected from the ultrasound generator.
Video available at:
https://www.youtube.com/watch?v=NXbGc6nAb5c
Disassembly
Release the permanently mounted standard/booster horn SH from the ultrasonic converter
UW. To do so, place the ultrasonic converter on a
fi
rm base (non-slip, if possible).
Place one sickle spanner on the ultrasonic converter and the other sickle spanner
on the standard/booster horn. To dismount, press the
fi
rst sickle spanner against the
fi
rm
base, and very
fi
rmly press the other sickle spanner in the opposite direction.
One
sickle spanner is part of the scope of delivery of one standard/booster horn.