Bamo Pooltest 3 Скачать руководство пользователя страница 8

 

msa190-10 

 

2010-12 

 

page 8 

 

 
 

Compliance 

 
The DPD method is a published standard method in the UK, the European Community, USA, Australia and in many other countries. The Palintest DPD 
free and total chlorine tests are approved by the USEPA as accepted  versions of the Standard Method 4500-C1-G (0 – 5 mg/L). 
 
The Palintest photometer series has been independently  tested and has earned the European CE Mark of Compliance for electro-magnetic compatibility 
(EMC). 
 
 
 
 

Sample Dilution and Collection 

 
When the test result is outside the concentration range  of the test, the photometer will display the '>' symbol.  In such cases it is necessary to dilute the 
pool water and repeat the test. A Palintest dilution tube (PT 512) is available to dilute samples accurately.  
 
If the result is close to the top of the scale ( e.g. chlorine above 4 mg/L on the Chlorine-Free/5 or Chlorine-Total/5 range, calcium hardness above 300 
mg/L and cyanuric acid above 150 mg/L) and a more accurate result is required, a dilution may be used to increase the sensitivity of the test.  
 
The following points should be observed when taking water samples from th e pool. Take pool water samples from below the water surface. Collect in a 
clean plastic bottle and fill to the neck so as to avoid unnecessary airspace.  
 
The free chlorine or bromine may drop during sample  storage. Other parameters such as pH and alkalin ity can also change. Samples should therefore 
be tested as soon as possible after sample collection. 
 
 
 
 

6 REORDER CODES and ACCESSORIES 

 

Code  

Reference  

Designation 

190 821  

DPD1 x250  

Pack of 250 tablets for testing acc. To DPD1  

190 823  

DPD3 x250  

Pack of 250 tablets for testing acc. To DPD3 

190 824 

isocyanuric x250 

Pack of 250 tablets for testing isocyanuric acid  

 

 

Содержание Pooltest 3

Страница 1: ...Pooltest 3 Handheld colorimeter For chlorine and isocyanuric acid tests INSTRUCTIONS MANUAL...

Страница 2: ...nknown colour in refere nce to known colours To avoid subjective measurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured...

Страница 3: ...ause inaccurate results It is imperative that the tubes and light chamber are cl ean and dry The glassware must be clean and defect free Scratches and abrasions will permanently affect the accuracy of...

Страница 4: ...d to dissolve the tablet and ensure any remaining particles have settled 3 Take photometer reading immediately as the result may drift on standing Ensure the display shows the test range that you desi...

Страница 5: ...d 1 Fill test tube with sample to the 10 mL mark 2 Add one PHENOL RED tablet crush and mix to dissolve 3 Take the photometer reading Reagents AP 130 Phenol Red Tablets Cyanuric Acid Range 2 to 200 mg...

Страница 6: ...unlikely event of malfunction These error messages are mainly designed to assist service staff in diagnosing instrument faults In the event of an error message appearing on the photometer display cont...

Страница 7: ...blet to your sample prior to adding your DPD tablet Non chlorine shock dosing treatments for swimming pools are widely used to support chlorine treated pools Chemicals containing monopotassium persulp...

Страница 8: ...close to the top of the scale e g chlorine above 4 mg L on the Chlorine Free 5 or Chlorine Total 5 range calcium hardness above 300 mg L and cyanuric acid above 150 mg L and a more accurate result is...

Отзывы: