3 TEST PROCEDURES 

 
 
The methodologies listed here are for use with the Pooltest 3 and Pooltest 6 instruments.   
Before attempting to do any tests, ensure you read ‘Blanks and Samples’ and the ‘Care and Maintenance’ sections in Section 2.  
 
 
 

To Select the Test 

 
Press the MENU key.  
The currently selected test will appear on the display:  
 
• Cl

5  

Chlorine   [Free and Total]  

(0.01 – 5 mg/L range) 

 
• pH  

pH  

(pH Units) 

 
• CNA  

Cyanuric Acid  

 
To change the test selected, press the MENU key  repeatedly until the required test appears on the display.  
The ranges in each test refer to the lower limit of detection to the highest value the unit can guarantee  confidence up to. 
 
 
 
 
 

Chlorine 
Range 
Cl

2

 5: 0.01 to 5 mg/L 

Wavelength – 530 nm 

 
 
 

Method – DPD 
 
 
 

Free Chlorine 
 

1 Rinse test tube with sample leaving two or three drops in the tube.  
 
2 Add one DPD No 1 tablet (or DPD-XF for high range instruments), crush the tablet with the crushing rod and then fill the test tube with sample to the 

10 ml mark.  

   Mix gently with the rod to dissolve the tablet and ensure any remaining particles have settled.  
 
3 Take photometer reading immediately as the  result may drift on standing. Ensure the display shows the test range that you desire.  
 
4 Retain test solution if Total Chlorine test required (see further on for more instructions). 
 
 
 

Total Chlorine 
 

1 Carry out this test on the solution remaining from the Free Chlorine test.  
 
2 If any non chlorine shock treatment chemicals have  been added to the pool, see addit ional notes further on 
 
3 Add one DPD No 3 tablet (or DPD-XT for high range instruments), crush and mix to dissolve.  
 
4 Stand for two minutes.  
 
5 Take photometer reading.  
   Ensure the display shows the test range that you desire.  
   To obtain COMBINED CHLORINE residual, subtract Free Chlorine result from Total Chlorine result: 
   i.e. Combined Chlorine = Total Chlorine - Free Chlorine  
   For notes on chlorine testing, see further on.  
 
 
 

Notes 

 
• If any non chlorine shock treatment chemicals have  been added to the pool, add one DPD Oxystop (AP 017)  tablet, crush and mix to dissolve prior to 

adding the DPD No 3 tablet to the test solution.  

   Stand for one minute before proceeding.  
   This will prevent a response caused by the shock treatment chemicals. 
 

Содержание Pooltest 3

Страница 1: ...Pooltest 3 Handheld colorimeter For chlorine and isocyanuric acid tests INSTRUCTIONS MANUAL...

Страница 2: ...nknown colour in refere nce to known colours To avoid subjective measurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured...

Страница 3: ...ause inaccurate results It is imperative that the tubes and light chamber are cl ean and dry The glassware must be clean and defect free Scratches and abrasions will permanently affect the accuracy of...

Страница 4: ...d to dissolve the tablet and ensure any remaining particles have settled 3 Take photometer reading immediately as the result may drift on standing Ensure the display shows the test range that you desi...

Страница 5: ...d 1 Fill test tube with sample to the 10 mL mark 2 Add one PHENOL RED tablet crush and mix to dissolve 3 Take the photometer reading Reagents AP 130 Phenol Red Tablets Cyanuric Acid Range 2 to 200 mg...

Страница 6: ...unlikely event of malfunction These error messages are mainly designed to assist service staff in diagnosing instrument faults In the event of an error message appearing on the photometer display cont...

Страница 7: ...blet to your sample prior to adding your DPD tablet Non chlorine shock dosing treatments for swimming pools are widely used to support chlorine treated pools Chemicals containing monopotassium persulp...

Страница 8: ...close to the top of the scale e g chlorine above 4 mg L on the Chlorine Free 5 or Chlorine Total 5 range calcium hardness above 300 mg L and cyanuric acid above 150 mg L and a more accurate result is...

Отзывы: