• DPD reacts with both chlorine and bromine. However, it is possible to determine bromine in the presence of chlorine, and to separate between bromine 
and chlorine residuals using a DPD methodology. For instructions on how to do this, please contact BAMO Mesures SAS.  
 
 
 

Reagents 

 
AP 011 - DPD No 1 Tablets  
 
 
 
 
 

pH 
Range: 6.5 to 8.4 
Wavelength: 575 nm 

 
 

Method – Phenol Red 

 
 
1 Fill test tube with sample to the 10 mL mark. 
 
2 Add one PHENOL RED tablet, crush and mix to  dissolve. 
 
3 Take the photometer reading. 
 

Reagents:

 AP 130 - Phenol Red Tablets 

 
 
 
 
 
 

Cyanuric Acid 
Range: 2 to 200 mg/L 
Wavelength 530 and 575 nm 

 
 

Method – Turbidity 

 
 
1 Fill test tube with sample to the 10 mL mark. 
 
2 Add one Cyanuric Acid tablet and allow to disintegrate  for at least two minutes. 
 
3 A cloudy solution indicates the presence of cyanuric acid. 
 
4 Crush any remaining undissolved tablet and mix to  ensure uniformity. 
 
5 Take the photometer reading. 
 

Reagents:

 AP 087 – Cyanuric Acid Tablets 

 
 

Содержание Pooltest 3

Страница 1: ...Pooltest 3 Handheld colorimeter For chlorine and isocyanuric acid tests INSTRUCTIONS MANUAL...

Страница 2: ...nknown colour in refere nce to known colours To avoid subjective measurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured...

Страница 3: ...ause inaccurate results It is imperative that the tubes and light chamber are cl ean and dry The glassware must be clean and defect free Scratches and abrasions will permanently affect the accuracy of...

Страница 4: ...d to dissolve the tablet and ensure any remaining particles have settled 3 Take photometer reading immediately as the result may drift on standing Ensure the display shows the test range that you desi...

Страница 5: ...d 1 Fill test tube with sample to the 10 mL mark 2 Add one PHENOL RED tablet crush and mix to dissolve 3 Take the photometer reading Reagents AP 130 Phenol Red Tablets Cyanuric Acid Range 2 to 200 mg...

Страница 6: ...unlikely event of malfunction These error messages are mainly designed to assist service staff in diagnosing instrument faults In the event of an error message appearing on the photometer display cont...

Страница 7: ...blet to your sample prior to adding your DPD tablet Non chlorine shock dosing treatments for swimming pools are widely used to support chlorine treated pools Chemicals containing monopotassium persulp...

Страница 8: ...close to the top of the scale e g chlorine above 4 mg L on the Chlorine Free 5 or Chlorine Total 5 range calcium hardness above 300 mg L and cyanuric acid above 150 mg L and a more accurate result is...

Отзывы: