background image

Page

13

450 M1 02 B

11-10-2007

Regulation

Gain

Ti(s)

Td(s)

PID

1,2 x T/t

2 x t

0,5 x t

PI

0,9 x T/t

3,3 x t

0

P

T/t

9999

0

ADJUST PID PARAMETERS 

In order to determinate the setup values for PID regulation, we recommend to use the Ziegler-Nichols open loop method

Proceed as following: 

- Connect a recorder to the analogic measurement output or write the reading measurement values for then to draw the graph f

(time)

- Switch on the PID regulation in MANUAL mode  
- Reach to and keep close the measurement value to the set point, adjusting the PID output 
- Apply on ΔCde a step of 10 % (for instance) on the analogic output (Cde).

Example:

if the value is 30,00 %, apply 40,00 %

- Note on the graph the corresponding timing. 
- Determinate on this graph both t ant T:

t = delay of response 
T = Time corresponding to the same variation in % of measurement (Δm) and the analogic output (ΔCde), Δm = ΔCde. 
This value may be found out on the slope. 

- Modify the PID parameters as following: 

REGUL PID

Output mA Oxygen 

Please see page 14

This operating mode allows a PID regulation with
an analogic output 0/20 or 4/20 mA

REGUL  

ON/OFF

To switch ON or OFF the regulation mode

REGUL  

AUTO/MANU

MANU=MANUAL to be able to check the relays output

SET VAL 

00,00 mg/l

Set point value

GAIN : 

0,000

Gain setup (see also ADJUST PID PARAMETERS)

T.i : 

0050

Sec

Integrant setup

Td : 

0012 

Sec

Derivative setup

ACTION:

DIRECT/REVERSE

OUTPUT 

4/20 mA 

/

0/20 mA

SAVING ?

L

L

L

L

L

L

L

L

L

L

L

L

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

VALID

MENU

+

To switch the PID regulation on stand-by, please input 24 V= 20 mA on terminals 5(+) and 6(-).

Содержание BAMOPHOX 450 E

Страница 1: ...amo fr Fax 33 01 34 10 16 05 E mail info bamo fr Dissolved oxygen monitoring BAMOPHOX 450 11 10 2007 450 M1 02 B MES 450 02 1 INSTRUCTION MANUAL INSTRUCTION MANUAL B B B BA A A AM M M MO O O OP P P PH...

Страница 2: ...3 WIRING 4 4 FRONT PANEL 6 PRESENTATION AND SCROLLING MENU 7 ABOUT BAMOPHOX 8 CONSULTATION MODIFICATION 8 MEASUREMENT PARAMETERS 9 ATMOSPHERIC PRESSURE 9 SALINITY 9 ADJUST THRESHOLD 1 10 ADJUST THRESH...

Страница 3: ...in 3 A 125 V AC 3 A 30 V DC and 105 evaluated charge for 3 A 125 V AC ON OFF Regulation Pulse time 0 9999 sec High and low proportional bandwidth high and low dead zones PID Regulation Proportionality...

Страница 4: ...Page 4 450 M1 02 B 11 10 2007 3 WIRING PANEL MOUNTING MODEL OPTION LOGGER RS 422 OPTION LOGGER access by the upper cover RS 422 WALL MOUNTING MODEL...

Страница 5: ...perature S1 threshold NO contact Measurement Temperature S3 threshold NO contact Measurement Temperature NOT USED Ground N Power supply N Neutral L Live L NC mA BAMOX 7200 sensor mA 1 2 3 4 5 6 7 8 9...

Страница 6: ...this key all parameters are saved previous data programmation will be overwrited If you are not sure of any modification do not press the VALID key To change parameters of data capture Num ric input...

Страница 7: ...CATION MEASUREMENT PARAMETERS ATMOSPHERIC PRESSURE SALINITY ADJUST ALARM 1 ADJUST ALARM 2 ADJUST ALARM 3 RELAY REGULATION REGUL PID OUTPUT mA Oxygen OUTPUT mA TEMP TEMPERATURE FORCED RELAY ADJUST ALAR...

Страница 8: ...tely the instrument acts within the configuration thresholds regulation analog outputs When PID regulation is activated the display shows the PID FORCED PID 00 00 mg l 0 000 one digit is flashing Modi...

Страница 9: ...00 00 C Example Real value 22 C Reference 20 C Entry 2 00 C SAVING ATMOSPHERIC PRESSURE P 1013 hPa The blinding digit has to be modified if necessary Confirm with Enter then save SAVING SALINITY g kg...

Страница 10: ...1 SAVING ADJUST ALARM 2 ADJUST ALARM 3 please see page 11 ALARM 2 ON OFF ALARM 2 MEASURE TEMP MEASURE Threshold against pH mV measured value TEMP Threshold against temperature measured value ALARM 2 L...

Страница 11: ...ntact closes when the value goes under the limit ON 00 00 mg l C To close the contact S3 at this value OFF 00 00 mg l C To open the contact S3 at this value DELAY UP ONI OFF Delay or no delay before t...

Страница 12: ...int T CYCLE 0000 SEC Cycle time Highest limit for HIGH PB 00 00 mg l proportional bandwith Lowest value LOW PB 00 00 mg l for dead zone Highest value H IGH DZ 00 00 mg l for dead zone Lowest value LOW...

Страница 13: ...e corresponding timing Determinate on this graph both t ant T t delay of response T Time corresponding to the same variation in of measurement m and the analogic output Cde m Cde This value may be fou...

Страница 14: ...mA TEMP TEMPERATURE HIGHER 0000 C Value corresponding to 20 00 mA LOWER 0000 C Value corresponding to 00 00 or 04 00 mA OUTPUT 4 20 mA 0 20 mA Output type SAVING L L L L L ENTER ENTER ENTER ENTER VALI...

Страница 15: ...or too small When Error message occurs press MENU to be back to CONFIGURATION Proceed to a new calibration sequence If the ERROR message appears again cancel the calibration sequence go back to measu...

Страница 16: ...LID MENU FORCED RELAY ADJUST ALARM ALARM 1 ON OFF ALARM 2 ON OFF Diagnostic mode to test the threshold configurations ALARM 3 ON OFF ALARM 4 ON OFF L L L L ENTER ENTER ENTER ENTER VALID MENU ADJUST AL...

Отзывы: