Page
8
450 M1 02 B
11-10-2007
ABOUT Bamophox
ABOUT BAMOPHOX
VERSION 2.04
SERIAL N°
20451 05
CONSULTATION / MODIFICATION
CONSULTATION
MODIFICATION
CODE ? 0000
CODE ? 5105
TIME : 30 mn
ENTER
ENTER
ENTER
ENTER
L
ENTER
ENTER
ENTER
MENU
+
L
L
From this mode MODIFICATION it is easy to return back to measurement for testing the relay outputs
and regulation mode.
Once in modification mode,
reach measurement display and press ENTER
FORCED MEASURE
00,00 mg/l +20°C
(one digit is flashing) Modify the value. Immediately the
instrument acts within the configuration (thresholds,
regulation, analog outputs ...).
When PID regulation is activated, the display shows the PID %
FORCED PID
00,00 mg/l 0,000%
(one digit is flashing) Modify the value. Immediately the
instrument acts within the configuration.
To test the analog output (mA) on PID mode: the PID should be active and in MANUAL mode.
Press ENTER to cancel the test mode and to go back to the measurement mode.
L
L
ENTER
ENTER
ENTER
ENTER
ENTER
L
L
Last 4 digits (of serial number) are the key code to
access the MODIFICATION menu.
When wrong code is entered, a message
"ERROR"
appears during 3 seconds.
After 30 minutes, the display returns automatically to the
measurement mode.