9.4

SETTING THE TEMPERATURE

Go to the menu TEMPERATURE

MEASURE : AUTO / MANUAL

AUTO : Temperature is displayed if a Pt 100 sensor is connected, confirm, then save.
MANUAL : Without Pt 100 sensor connected, enter the temperature manually.
Confirm

FLUID T. +000.0 °C

In the MANUAL mode, enter the temperature of the liquid, then confirm.

SAVE ?

To record the settings, press the icon SAVE

9.5

mA OUTPUT SETTINGS FOR THE MEASUREMENT

Go to the menu OUTPUT mA TURB

HIGHER 2000

Enter the measured value for which the output is 20 mA, then confirm.

LOWER 0000

Enter the measured value for which the output is 0 or 4 mA, then confirm

OUTPUT 4/20 mA or 0/20 mA

Select the output signal type, then confirm

SAVE ?

To record the settings, press the icon SAVE

9.6

mA OUTPUT SETTINGS FOR THE TEMPERATURE

Go to the menu OUTPUT mA TEMP.

160.0 °C

Enter the temperature value for which the output is 20 mA, then confirm.

LOWER +000.0 °C

Enter the temperature value for which the output is 0 or 4 mA, then confirm.

OUTPUT 4/20 mA or 0/20 mA

Select the output signal type, then confirm

SAVE ?

To record the settings, press the icon SAVE

9.7

SIMULATION OF RELAYS

With the menu FORCED RELAY, it is possible to manually test (force) relays S1, S2, S3 and S4.

Last one, S4, is N.O. by default. The test begins by relay S1.
The respective relay can be switched from OFF (open) to ON (closed).
Confirm to go to the next relay, then to go back to previous menu.

9.8

SIMULATION OF MEASUREMENT

Go to the menu FORCED MEASURE

0.000

Simulation at this measured value.
First displayed value is the current measurement.
Confirm, then enter the value you wish to test.

Note: The value is immediately considered by the monitor and actuates thresholds, 4-20mA outputs, etc.

TUR

07-06-2021

M-436.02-EN-AE

436-02

/11

Turbidity monitor

BAMOPHAR 436

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Содержание BAMOPHAR 436

Страница 1: ...MANUAL TUR 07 06 2021 M 436 02 EN AE 436 02 1 Turbidity monitor BAMOPHAR 436 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www ba...

Страница 2: ...9 CONSULTATION MODIFICATION 9 10 SETTINGS 9 1 10 SETTING THE MEASUREMENT 9 2 10 THRESHOLDS SETTING FOR S1 S2 S3 9 3 10 ASSIGNING S3 TO AN EXTERNAL SIGNAL 9 4 11 SETTING THE TEMPERATURE 9 5 11 mA OUTP...

Страница 3: ...contacts potential free S1 threshold for measurement or temperature N O S2 threshold for measurement or temperature N O S3 threshold for measurement or temperature N O or assigned to an external conta...

Страница 4: ...wall mount version accessible after removing the top cover Measurement signal input 4 20 mA TUR 07 06 2021 M 436 02 EN AE 436 02 4 Turbidity monitor BAMOPHAR 436 22 Rue de la Voie des Bans Z I de la g...

Страница 5: ...measurement 17 S4 18 Relay S2 N O contact for measurement signal 19 S2 20 Relay 1 N O contact for measurement 21 S1 22 Relay 3 N O contact for measurement 23 S3 24 Not used 25 Main power supply 26 PE...

Страница 6: ...ITTERS TUR 07 06 2021 M 436 02 EN AE 436 02 6 Turbidity monitor BAMOPHAR 436 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www bam...

Страница 7: ...time Only for data logger versions Display settings Measurement details Contact symbol appears if relay is set up Switching status of relays S1 S2 S3 S4 Designation of each measuring systems Measurem...

Страница 8: ...age selection MENU Access to monitoring parameters INFO Access to serial number and version of BAMOPHAR PADLOCK Open MODIFICATION MODE Closed CONSULTATION MODE Query mode RETURN KEY Return to previous...

Страница 9: ...the operator to check out all working parameters This mode is represented by the closed padlock icon To change the settings of the BAMOPHAR you need to enter in the MODIFICATION mode This mode is pro...

Страница 10: ...can be set for signal MEASUREMENT or TEMPERATURE MEASUREMENT Threshold dedicated to the monitoring of turbidity signal TEMPERATURE Threshold dedicated to the monitoring of temperature signal Select y...

Страница 11: ...settings press the icon SAVE 9 6 mA OUTPUT SETTINGS FOR THE TEMPERATURE Go to the menu OUTPUT mA TEMP HIGHER 160 0 C Enter the temperature value for which the output is 20 mA then confirm LOWER 000 0...

Отзывы: