1.
DESCRIPTION
BAMOPHAR 436 is designed for turbidity measurements for control and monitoring of water.
Connected to a transmitter TURBICUBE or TRUBOMAT GAB for in-line measurement or to a TURBISENS for measurement in immersion,
BAMOPHAR will display and transmit 2 analogue signals, as 4-20 mA, for temperature and turbidity.
BAMOPHAR displays on a color touch-sensitive screen a multilingual friendly using menu.
The reading is easy for measurement, temperature and relay status.
It displays a menu with all parameters for configuration of analogue outputs and thresholds.
For a safe start-up a test menu allow a simulation of operation actuating all outputs and thresholds.
2.
TECHNICAL FEATURES
End-user interface
Color touch screen 4.3'', resolution 480x272 pixels
Display of measurements, menus, temperature, relay status
Configuration - Keyword protected
Measuring scales
To set up according the turbidity transmitter in use (from 0.001 to 9999 FNU, FAU, mg/l, g/l)
Input signal
For turbidity transmitter with 4-20 mA output signal.
Temperature display
When a Pt 100 Ω (3-wire) is connected: range 0 ... 100 °C
Or manual set up
Relay outputs
4 contacts, potential free
S1, threshold for measurement or temperature (N.O.)
S2, threshold for measurement or temperature (N.O.)
S3, threshold for measurement or temperature (N.O.) or assigned to an external contact
S4, contact as alarm function (fault on Pt 100 loop) or over-riding measurement
Initial contact resistance
100 mΩ max. (voltage drop 6 V DC 1 A)
Switching power
831 VA AC / 3 A / 277 V AC
90 W / 3 A / 30 V DC
Switching capacity (min.)
100 mA, 5 V DC (depends of frequency, ambient conditions)
Mechanical lifetime (min.)
5 x 10⁶ operations (180 op./min)
Electrical lifetime (min.)
2 x 10⁵ (at 20 op./min) for 3 A 125 V AC, 3 A 30 V DC - 10⁵ (estimated load) for 3 A 125 V AC
Output measurement signal
0/4 - 20 mA (max. 600 Ω), directly proportional to measurement
Power supply
230 V, 50/60 Hz (other on request); Consumption 10 VA
Models
Panel mounting, 96x144 mm, Front IP65, screw terminal IP40
Wall mounting, IP65, cable glands, screw terminals
OPTION (RS 422 + Logger)
Interface
RS422 output, J-BUS link - Binary slave mode - 2400 to 9600 bauds
Data Logger: Automatic record of cycle average measurement
150 000 records max. on memory card
3.
DIMENSIONS
TUR
07-06-2021
M-436.02-EN-AE
436-02
/3
Turbidity monitor
BAMOPHAR 436