background image

1.

DESCRIPTION

BAMOPHAR 436 is designed for turbidity measurements for control and monitoring of water.
Connected to a transmitter TURBICUBE or TRUBOMAT GAB for in-line measurement or to a TURBISENS for measurement in immersion,
BAMOPHAR will display and transmit 2 analogue signals, as 4-20 mA, for temperature and turbidity.
BAMOPHAR displays on a color touch-sensitive screen a multilingual friendly using menu.

The reading is easy for measurement, temperature and relay status.
It displays a menu with all parameters for configuration of analogue outputs and thresholds.
For a safe start-up a test menu allow a simulation of operation actuating all outputs and thresholds.

2.

TECHNICAL FEATURES

End-user interface

Color touch screen 4.3'', resolution 480x272 pixels
Display of measurements, menus, temperature, relay status

Configuration - Keyword protected

Measuring scales

To set up according the turbidity transmitter in use (from 0.001 to 9999 FNU, FAU, mg/l, g/l)

Input signal

For turbidity transmitter with 4-20 mA output signal.

Temperature display

When a Pt 100 Ω (3-wire) is connected: range 0 ... 100 °C
Or manual set up

Relay outputs

4 contacts, potential free
S1, threshold for measurement or temperature (N.O.)
S2, threshold for measurement or temperature (N.O.)

S3, threshold for measurement or temperature (N.O.) or assigned to an external contact
S4, contact as alarm function (fault on Pt 100 loop) or over-riding measurement

Initial contact resistance

100 mΩ max. (voltage drop 6 V DC 1 A)

Switching power

831 VA AC / 3 A / 277 V AC

90 W / 3 A / 30 V DC

Switching capacity (min.)

100 mA, 5 V DC (depends of frequency, ambient conditions)

Mechanical lifetime (min.)

5 x 10⁶ operations (180 op./min)

Electrical lifetime (min.)

2 x 10⁵ (at 20 op./min) for 3 A 125 V AC, 3 A 30 V DC - 10⁵ (estimated load) for 3 A 125 V AC

Output measurement signal

0/4 - 20 mA (max. 600 Ω), directly proportional to measurement

Power supply

230 V, 50/60 Hz (other on request); Consumption 10 VA

Models

Panel mounting, 96x144 mm, Front IP65, screw terminal IP40

Wall mounting, IP65, cable glands, screw terminals

OPTION (RS 422 + Logger)

Interface

RS422 output, J-BUS link - Binary slave mode - 2400 to 9600 bauds

Data Logger: Automatic record of cycle average measurement
150 000 records max. on memory card

3.

DIMENSIONS

TUR

07-06-2021

M-436.02-EN-AE

436-02

/3

Turbidity monitor

BAMOPHAR 436

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Содержание BAMOPHAR 436

Страница 1: ...MANUAL TUR 07 06 2021 M 436 02 EN AE 436 02 1 Turbidity monitor BAMOPHAR 436 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www ba...

Страница 2: ...9 CONSULTATION MODIFICATION 9 10 SETTINGS 9 1 10 SETTING THE MEASUREMENT 9 2 10 THRESHOLDS SETTING FOR S1 S2 S3 9 3 10 ASSIGNING S3 TO AN EXTERNAL SIGNAL 9 4 11 SETTING THE TEMPERATURE 9 5 11 mA OUTP...

Страница 3: ...contacts potential free S1 threshold for measurement or temperature N O S2 threshold for measurement or temperature N O S3 threshold for measurement or temperature N O or assigned to an external conta...

Страница 4: ...wall mount version accessible after removing the top cover Measurement signal input 4 20 mA TUR 07 06 2021 M 436 02 EN AE 436 02 4 Turbidity monitor BAMOPHAR 436 22 Rue de la Voie des Bans Z I de la g...

Страница 5: ...measurement 17 S4 18 Relay S2 N O contact for measurement signal 19 S2 20 Relay 1 N O contact for measurement 21 S1 22 Relay 3 N O contact for measurement 23 S3 24 Not used 25 Main power supply 26 PE...

Страница 6: ...ITTERS TUR 07 06 2021 M 436 02 EN AE 436 02 6 Turbidity monitor BAMOPHAR 436 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www bam...

Страница 7: ...time Only for data logger versions Display settings Measurement details Contact symbol appears if relay is set up Switching status of relays S1 S2 S3 S4 Designation of each measuring systems Measurem...

Страница 8: ...age selection MENU Access to monitoring parameters INFO Access to serial number and version of BAMOPHAR PADLOCK Open MODIFICATION MODE Closed CONSULTATION MODE Query mode RETURN KEY Return to previous...

Страница 9: ...the operator to check out all working parameters This mode is represented by the closed padlock icon To change the settings of the BAMOPHAR you need to enter in the MODIFICATION mode This mode is pro...

Страница 10: ...can be set for signal MEASUREMENT or TEMPERATURE MEASUREMENT Threshold dedicated to the monitoring of turbidity signal TEMPERATURE Threshold dedicated to the monitoring of temperature signal Select y...

Страница 11: ...settings press the icon SAVE 9 6 mA OUTPUT SETTINGS FOR THE TEMPERATURE Go to the menu OUTPUT mA TEMP HIGHER 160 0 C Enter the temperature value for which the output is 20 mA then confirm LOWER 000 0...

Отзывы: