Technical data
Electric data
Operating voltage U
b
18…30 V DC
No-load current I0 max
≤ 30 mA
Interface
IO-Link V1.1
Polarity reversal protected
yes
Short circuit protected
yes
Switching frequency
200 Hz
Accuracy per IEC 60770
≤ ±0.5% FSO BFSL
Temperature error
≤ ±0.5% FSO/10 K
Mechanical data
Housing material
Stainless steel 1.4301
Measuring cell material
Ceramic with FKM
seal (internal)
Sealing ring material (only
versions with external G-thread)
Fluoroelastomer
Connector
M12, 4-pole
Process connection material
Stainless steel 1.4301
Process connection
G 1/4"
1/4" NPT
Ambient conditions
Ambient temperature
–25…+85 °C
Medium temperature
–40…+125 °C
Degree of protection as per IEC
60529
IP67 when threaded
together
Your pressure transmitter type
The part label contains the exact part number and key
technical data for uniquely identifying the device.
Rel. nominal
pressure
Overload
pressure
Burst pressure
Sensor –1…2 bar
5 bar
35 bar
Sensor –1…10 bar
20 bar
10 bar
Sensor 0…2 bar
5 bar
10 bar
Sensor 0…5 bar
12 bar
15 bar
Sensor 0…10 bar
20 bar
35 bar
Sensor 0…20 bar
50 bar
70 bar
Sensor 0…50 bar
120 bar
150 bar
Sensor 0…100 bar
200 bar
300 bar
Sensor 0…250 bar
400 bar
750 bar
Sensor 0…400 bar
650 bar
1000 bar
Sensor 0…600 bar
1000 bar
1570 bar
Part label
The part label contains the exact part number and key
technical data for uniquely identifying the device.
Part label (example)
1)
Pin 1:
Pin 2:
Pin 3:
Pin 4:
2136
D
E S#
11124561
BSP B250-FV004-D06S1A-S4
BSP00TF
Balluff GmbH
Schurwaldstrasse 9 - D-73765 Neuhausen a.d.F.
Phone +49 7158 173-0
+24 Vdc
n.c.
0V GND
PNP (Out) /
0...250 bar
0...25 MPa
0...3625 psi
4XV7
PROC. CONT. EQ.
1)
2)
3)
4)
Ordering code
2)
Type
3)
Electrical connections
4)
Nominal pressure