Page 50 of 61
CLEANING AND SERVICE REQUIREMENTS
Service and Cleaning Overview
To maintain optimal performance of the InvivO
2
workstation it must be serviced at regular
intervals. Lists the servicing requirements, intervals and persons capable of performing
the service are detailed in Table 9.
Important notice:
Your Baker Ruskinn Workstation includes a 2 year warranty. This is only valid if the first
year service is carried out in accordance with table 6 below.
Action
Frequency
By
Clean workstation
After each use
End User
Empty waste port bag
Weekly
End User
Fill humidifier tanks
Weekly
End User
Calibrate oxygen sensors
Monthly
End User
Deep clean workstation
3-6 months
End User
Replace detox sachets
Annually
End User/ Qualified service
engineer
Replace catalyst sachets (if
used)
Annually
End User/ Qualified service
engineer
End User service
1
st
and 3
rd
Years
End User
Preventative Maintenance
Service
2
nd
and 4
th
Years
Qualified service engineer
Table 9: Cleaning and Service Details
To order end user service kits or to arrange a service engineer visit contact your local
distributor. For a service engineer visit please ensure that the unit is in a safe state and
decontaminated if required.
To arrange a service contract contact your local distributor.