Bugbox M
UM-20 R4.0
Page 43 of 47
Bugbox M User Manual
Interlock problems
Figure 33 gives a list of common interlock problems and solutions;
Figure 33: Interlock problems and solutions
Problem
Cause
Solution
The interlock inner
door will not close
The interlock inner
door is blocked
The interlock inner
door is stuck
Check for obstructions to closing the interlock
inner door and remove them
Spillages can cause the interlock inner door to
stick. Clean the interlock inner door. See section
7.1.2 / 7.1.3 for more details
Apply a small amount of port Lube Talc to the
inner interlock door seal.
The interlock outer
door will not close
The interlock outer
door is obstructed
The interlock outer
door catch has broken
Check for obstructions to closing the interlock
outer door and remove them
Contact your local distributor
Gas mixer and gas consumption problems
The Bugbox M workstation is fitted with a gas demand indicator (item 7 in Figure 8). When the gas
demand indicator is illuminated, the pressure is low in the workstation chamber. A shot of gas will be
injected into the workstation to raise the pressure in the workstation chamber. In normal operation,
the gas demand indicator should not illuminate more than once every 20 minutes when the
workstation is at rest. At rest is defined as the gasses are within deadband and glove ports and SPES
are closed. Figure 34 gives a list of common problems related to the gas mixer and gas consumption;
Figure 34: ICO
2
N
2
IC workstation related and gas consumption common problems
Problem
Potential Causes
Solution
The gas demand indicator
illuminates more than
once every 20 minutes
whilst at rest/ high gas
consumption
The Ezee Sleeves are not
correctly installed
The glove port cover is not
correctly closed
The cuff is not tight
around the users wrist
The humidifier tank is
empty
A gas cylinder is empty
Check the Ezee Sleeves are correctly installed. See
section 7.2.2, for more details
Check the glove port cover is in place and closed
correctly. Reclose the glove port cover if required
Ensure that the correct size Ezee Sleeve is used.
Ensure that the cuff is pulled tight around the
wrist. Roll up sleeves of clothing and remove
jewellery
Refill the humidifier tank. See section 7.2.1, for
more details
Check the ICO
2
N
2
IC gas mixer screen to see which
gas cylinder is empty. An empty cylinder will be
indicated by a low pressure status. Replace the
gas cylinder with a full cylinder