Replacing detox sachets
Detox sachets are supplied with the Bugbox M workstation. The detox sachets adsorb volatile organic
compounds, improving the air quality within the workstation chamber. Detox sachets need to be
replaced annually. See section 7.4.2 for ordering details.
To replace the detox sachets;
Remove any samples to an alternative storage facility.
Return the ICO
2
N
2
IC to the main menu.
Switch off Bugbox M and disconnect from the mains power supply.
Remove the Ezee Sleeves (see section 7.2.2 for more information).
Undo the glove port covers and place on the rear shelf.
Lift the floor tray at the front to expose the detox sachets. Figure 26 shows the location of the
sachet holders;
Figure 26: Sachet holder location
1.
Floor tray.
2.
Sachet holder. The second sachet holder is located behind and to the
left of the first sachet holder.
3.
Floor tray locator.
Slide the sachets forwards to remove them from their holders.
Remove the sachets via the glove ports.
Lower the floor tray, ensuring that the floor tray leg has located in the floor tray locator (item
3 in Figure 26).
Remove the new sachets from their packaging.
Whilst lifting the floor tray at the front, slide the sachets into the sachet holder.
1
2
3