Page
46
of
55
Cleaning procedure
–
deep clean
To deep clean the workstation;
Preparing the workstation
•
Remove all cells/ samples to an alternative storage facility.
•
Switch the workstation off at the mains and remove the plug from the mains.
•
Remove the Ezee Sleeves
.
•
Remove any other equipment installed in the workstation chamber.
Cleaning the workstation chamber
•
The items inside the workstation chamber that require cleaning are;
o
The floor tray.
o
The ceiling panel.
o
The shelf panel.
o
The left hand wall.
o
The right hand wall.
o
The interlock inner door.
o
The glove port covers.
o
If fitted, the inside of the SPES.
•
For all workstation components, wipe with a paper towel soaked in an appropriate cleaning
agent. Take care not to get cleaning agents on the fans (below the shelf panel) and if fitted,
the plug socket
•
Wipe dry using paper towels.
Cleaning the interlock chamber;
•
The items inside the interlock chamber that require cleaning are;
o
The floor.
o
The interlock outer door
o
The Petri dish holder
o
The inside wall
o
The interlock inner door.
•
For all interlock components, wipe with a paper towel soaked in an appropriate cleaning
agent.
•
Wipe dry using paper towels.
Reinstalling workstation components
•
Reinstall any equipment removed from the workstation chamber.
•
Reinstall the Ezee Sleeves.