Service Requirements
Electrical Supply Requirements
The workstation must be connected to a mains power supply. A power cord is supplied to connect
the workstation to the mains supply. If an alternative power cord is used must be rated
appropriately for the power requirements of the workstation, refer to Table 2. The workstation
must be connected to a protective earth.
To ensure safe operation of the workstation, it must be connected to a supply of the correct
voltage and frequency as stated on the rating label shown at the rear of the unit. The mains
supply voltage fluctuations must not /- 10% of the nominal mains voltage. The input
voltage and frequency for Bugbox Ax workstations are:
Voltage Range
Frequency
Nominal Power
220V to 240V AC 50/60Hz
560W
100V to 120 V AC
50/60Hz
430W
Table 2: Electrical Service Requirements
Please note that the applicable ratings for the Workstation configuration are detailed the
labelling at the rear of the product as shown in Figure 2.
Figure 2: Bugbox Ax Rear Connections
1.
Gas Supplies
2.
Residual Current Circuit Breaker (RCCB)
3.
Mains Input Connector (C19)
4.
Footswitch Connector
5.
Mains Input Fuses
6.
Earth Stud
2
1
5
4
3
6