3-30
CL 5
Creating and Erasing Custom Setup Files
Setting Up the CL 5
alarm is violated. Note that the inputted alarm value
will correspond to a thickness except when the VELO-
CITY measurement option is installed and activated.
In this case the alarm settings will represent velocity
values. To input a minimum or maximum alarm:
Step 1:
Press to access the Configuration display
(shown in Figure 3-3).
Step 2:
Press or to select
MINIMUM ALARM
or
MAXIMUM ALARM
Step 3:
When the desired parameter is highlighted,
press
to activate it.
Step 4:
Press or to change the selected posi-
tion in the alarm thickness and or to adjust the
value of the selected position. To switch the alarm
setting to OFF from any other value, simultaneously
press an .
Step 5:
Press when the alarm thickness is set to
the desired value.
3.5 Creating and Erasing
Custom Setup Files
After loading a Standard Setup and performing the
necessary adjustments to optimize performance for
a given application, the instrument settings can be
stored and named as a Custom Setup file. The instru-
ment is capable of storing up to 5 Custom Setup files,
which can be erased at any time
(Figure 3-5).
Custom Setups can be created from any CL 5 key-
pad. Custom Setups created using a base-model
instrument include the following settings:
• User-assigned Custom Setup file name
• VELOCITY
• MINIMUM ALARM
• MAXIMUM ALARM
Содержание Waygate Technologies CL 5
Страница 1: ...Krautkr mer CL 5 Operating Manual 021 002 296 rev 04...
Страница 2: ...CL 5 1 2 General Information 1...
Страница 10: ...CL 5 2 10 Understanding the Keypad Menu System and Displays 2...
Страница 20: ...CL 5 3 20 Setting Up the CL 5 3...
Страница 24: ...3 24 CL 5 Configuring the Instrument s Display Setting Up the CL 5 FIGURE 3 2 Selecting a Setup File...
Страница 26: ...3 26 CL 5 Configuring the Instrument s Display Setting Up the CL 5 FIGURE 3 3 Changing Parameter Settings...
Страница 35: ...CL 5 3 35 Setting Up the CL 5 Locking and Unlocking Instrument Controls THIS PAGE LEFT INTENTIONALLY BLANK...
Страница 36: ...CL 5 4 36 Measuring Thickness 4...
Страница 49: ...CL 5 5 49 Using the Optional Data Recorder 5...
Страница 56: ...5 56 CL 5 Printing a Report Using the Optional Data Recorder THIS PAGE LEFT INTENTIONALLY BLANK...
Страница 57: ...CL 5 6 57 I O Features 6...
Страница 59: ...CL 5 6 59 Transmitting Thickness Data Through an External Device I O Features...
Страница 69: ...CL 5 7 69 Specifications 7...
Страница 76: ...7 76 CL 5 Specifications CL 5 Probe Transducer Specifications THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Страница 77: ...CL 5 8 77 Maintenance 8...
Страница 79: ...CL 5 9 79 Appendix 9...
Страница 84: ...CL 5 10 84 Index 10...